A new stand-alone cervical anterior interbody fusion device: biomechanical comparison with established anterior cervical fixation devices

Matti Scholz, Phillip M Reyes, Philipp Schleicher, Anna G U Sawa, Seungwon Baek, Frank Kandziora, Frederick F Marciano, Neil R Crawford
Spine 2009 January 15, 34 (2): 156-60
19139665

STUDY DESIGN: A new anchored spacer-a low-profile cervical interbody fusion cage with integrated anterior fixation-was compared biomechanically to established anterior cervical devices.

OBJECTIVE: To evaluate the fixation properties of the new stand-alone device and compare these properties with established fixation methods. The hypothesis is that the new device will provide stability comparable to that provided by an anterior cervical cage when supplemented with an anterior plate.

SUMMARY OF BACKGROUND DATA: It is accepted that the use of anterior cervical plating increases the chance of achieving a solid fusion. However, its use may be associated with an increase in operation time and a higher postoperative morbidity caused by a larger anterior approach and disruption of the anterior musculature. This dilemma has led to the development of a new, low profile stand-alone cervical anterior cage device with integrated screw fixation.

METHODS: Twenty-four human cadaveric C4-C7 cervical spines were loaded nondestructively with pure moments in a nonconstraining testing apparatus to induce flexion, extension, lateral bending, and axial rotation while angular motion was measured optoelectronically. The specimens were tested: 1. Intact (N = 24). 2. After discectomy and anterior stabilization. a. Interbody cage + locking plate (N = 8). b. Interbody cage + dynamic plate (N = 8). c. Anchored spacer (N = 8). 3. After ventral plate removal of group 2a and 2b (N = 16).

RESULTS: All fixation techniques decreased range of motion (ROM) and lax zone (LZ) (P < 0.05) in all test modes compared with the intact motion segment and cage-only group. There were no significant differences between the anchored spacer and cage + locking plate or cage + dynamic plate.

CONCLUSION: The anchored spacer provided a similar biomechanical stability to that of the established anterior fusion technique using an anterior plate plus cage and has a potentially lower perioperative and postoperative morbidity. These results support progression to clinical trials using the cervical anchored spacer as a stand-alone implant.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
19139665
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"