Improved renal function and blood pressure control following renal artery angioplasty: the renal artery angioplasty in patients with renal insufficiency and hypertension using a dedicated renal stent device study (PRECISION).
EuroIntervention 2008 August
AIMS: To evaluate the technical performance of a dedicated renal stent device and the clinical outcome.
METHODS AND RESULTS: Fifty patients with 55 renal artery stenoses (RAS) >70% (66 +/- 12 years, 58% male) were included in this non-randomised, prospective, multicentre registry. Primary endpoint was the primary patency rate at one year defined as > or =70% as determined by duplex ultrasound. Major secondary endpoints were procedural success, 30 days MACE rate, the impact of the intervention on renal function, blood pressure control, and on B-type natriuretic peptide (BNP) level. Procedural success rate was 100% and 30 days MACE rate was 0%. Restenosis rate (primary endpoint) and target lesion revascularisation rate after 12 months were 3.5% and 1.8%, respectively. After one year estimated glomerular filtration rate increased from 51 +/- 26 ml/min to 61 +/- 28 ml/min (P=0.004). Mean ambulatory blood pressure was reduced from 102 +/- 14 mmHg to 93 +/- 9 mmHg (P=0.001). Mean daily dose of antihypertensive drugs decreased from 3.0 +/- 1.7 to 2.7 +/- 1.4 (P=0.09). Mean BNP decreased from 251 +/- 282 pg/ml to 188 +/- 219 pg/ml (P=0.046) before discharge.
CONCLUSIONS: Technical outcome of the tested device is favourable. The impact of the stent revascularisation on renal function and blood pressure control was promising.
METHODS AND RESULTS: Fifty patients with 55 renal artery stenoses (RAS) >70% (66 +/- 12 years, 58% male) were included in this non-randomised, prospective, multicentre registry. Primary endpoint was the primary patency rate at one year defined as > or =70% as determined by duplex ultrasound. Major secondary endpoints were procedural success, 30 days MACE rate, the impact of the intervention on renal function, blood pressure control, and on B-type natriuretic peptide (BNP) level. Procedural success rate was 100% and 30 days MACE rate was 0%. Restenosis rate (primary endpoint) and target lesion revascularisation rate after 12 months were 3.5% and 1.8%, respectively. After one year estimated glomerular filtration rate increased from 51 +/- 26 ml/min to 61 +/- 28 ml/min (P=0.004). Mean ambulatory blood pressure was reduced from 102 +/- 14 mmHg to 93 +/- 9 mmHg (P=0.001). Mean daily dose of antihypertensive drugs decreased from 3.0 +/- 1.7 to 2.7 +/- 1.4 (P=0.09). Mean BNP decreased from 251 +/- 282 pg/ml to 188 +/- 219 pg/ml (P=0.046) before discharge.
CONCLUSIONS: Technical outcome of the tested device is favourable. The impact of the stent revascularisation on renal function and blood pressure control was promising.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app