Evaluation of an uncalibrated arterial pulse contour cardiac output monitoring system in cirrhotic patients undergoing liver surgery

G Biancofiore, L A H Critchley, A Lee, L Bindi, M Bisà, M Esposito, L Meacci, R Mozzo, P DeSimone, L Urbani, F Filipponi
British Journal of Anaesthesia 2009, 102 (1): 47-54
19059920

BACKGROUND: The pulmonary artery catheter is invasive and may cause serious complications. A safe method of cardiac output (CO) measurement is needed. We have assessed the accuracy and reliability of a recently marketed self-calibrating arterial pulse contour CO monitoring system (FloTrac/Vigileo) in end-stage liver failure patients undergoing liver transplant. The pattern of alterations known as cirrhotic cardiomyopathy, and the transplant procedure itself, provided an evaluation under varying clinical conditions.

METHODS: The cardiac index was measured simultaneously by thermodilution (CI(TD): mean of four readings) using a pulmonary artery catheter and pulse contour analysis (CI(V): mean value computed by the FloTrac/Vigileo over the same time period). Readings were made at 10 time-points during liver transplant surgery (T1-T5) and on the intensive care unit (T6-T10). CI(V) was computed using the latest Vigileo software version 01.10.

RESULTS: A total of 290 paired readings from 29 patients were collected. Mean (SD) CI(TD) was 5.2 (1.3) and CI(V) was 3.9 (0.9) litre min(-1) m(-2), with a corrected for repeated measures bias between readings of 1.3 (0.2) litre min(-1) m(-2) and 95% limits of agreement of -1.5 (0.2) to 4.1 (0.3) litre min(-1) m(-2). The percentage error (2SD(Bias)/meanCI(TD)) was 54%, which exceeded a 30% limit of acceptance. Low peripheral resistance and increasing bias were related (r=0.69; P<0.001). The Vigileo system failed to reliably trend CI data, with a concordance compared with thermodilution below an acceptable level (at best 68% of sequential readings).

CONCLUSIONS: In cirrhotic patients with hyperdynamic circulation, the Vigileo system showed a degree of error and unreliability higher than that considered acceptable for clinical purposes.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
19059920
×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"