Comparison of myocardial perfusion 82Rb PET performed with CT- and transmission CT-based attenuation correction.

UNLABELLED: CT-based attenuation correction (AC) for myocardial perfusion PET studies is challenging because of respiratory motion. Our study aimed to compare the transmission CT (TCT)-based and CT-based AC for myocardial perfusion PET/CT images with a direct semiquantitative approach comparing differences in segmental count distribution.

METHODS: Stress and rest (82)Rb PET scans from 54 consecutive patients acquired on a PET/CT scanner with dual CT-based and TCT-based AC were considered. TCT- and CT-based AC images were automatically registered to each other, and direct voxel-based and American Heart Association segment-based estimation of positive and negative changes between these scans was performed. Additionally, visual quality control (QC) of CT map alignment with PET emission data was performed by 2 expert observers, and studies with significant (>/=5 mm) misalignment were reprocessed with corrected CT alignment.

RESULTS: We used the 17-segment American Heart Association model for TCT-to-CT regional change analysis in all patients and found that 4 segments on rest and 4 segments on stress scans differed more than 3% between CT- and TCT-corrected images for studies without significant misalignments (<5 mm); only 1 differed by more than 5%. In cases with significant misalignment of greater than or equal to 3% TCT-CT AC, changes were observed on 14 rest and 10 stress segments; after alignment, these differences were still seen in 13 rest segments and 11 stress segments. Visual QC revealed that 46% of rest and 54% of stress PET scans were misaligned by greater than or equal to 5 mm with the CT maps acquired during normal breathing. The range of the reported PET/CT misalignment was 0-15 mm in x, 0-16 mm in y, and 0-20 mm in z directions. The overall agreement in visual QC of PET/CT alignment between the observers was 72.2%

CONCLUSION: There are significant differences between TCT and CT AC applied to cardiac PET/CT studies, which remain after alignment of CT maps to emission data.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app