Journal Article
Research Support, N.I.H., Extramural
Research Support, Non-U.S. Gov't
Add like
Add dislike
Add to saved papers

Prostate intrafraction motion evaluation using kV fluoroscopy during treatment delivery: a feasibility and accuracy study.

Margin reduction for prostate radiotherapy is limited by uncertainty in prostate localization during treatment. We investigated the feasibility and accuracy of measuring prostate intrafraction motion using kV fluoroscopy performed simultaneously with radiotherapy. Three gold coils used for target localization were implanted into the patient's prostate gland before undergoing hypofractionated online image-guided step-and-shoot intensity modulated radiation therapy (IMRT) on an Elekta Synergy linear accelerator. At each fraction, the patient was aligned using a cone-beam computed tomography (CBCT), after which the IMRT treatment delivery and fluoroscopy were performed simultaneously. In addition, a post-treatment CBCT was acquired with the patient still on the table. To measure the intrafraction motion, we developed an algorithm to register the fluoroscopy images to a reference image derived from the post-treatment CBCT, and we estimated coil motion in three-dimensional (3D) space by combining information from registrations at different gantry angles. We also detected the MV beam turning on and off using MV scatter incident in the same fluoroscopy images, and used this information to synchronize our intrafraction evaluation with the treatment delivery. In addition, we assessed the following: the method to synchronize with treatment delivery, the dose from kV imaging, the accuracy of the localization, and the error propagated into the 3D localization from motion between fluoroscopy acquisitions. With 0.16 mAs/frame and a bowtie filter implemented, the coils could be localized with the gantry at both 0 degrees and 270 degrees with the MV beam off, and at 270 degrees with the MV beam on when multiple fluoroscopy frames were averaged. The localization in two-dimensions for phantom and patient measurements was performed with submillimeter accuracy. After backprojection into 3D the patient localization error was (-0.04 +/- 0.30) mm, (0.09 +/- 0.36)mm, and (0.03 +/- 0.68)mm in the right-left (RL), anterior-posterior (AP), and superior-inferior (SI) axes, respectively. Simulations showed that while oscillating (stationary) motion cannot be effectively represented in 3D, linearly drifting (nonstationary) motion is detectable with good accuracy. These results show that measuring prostate intrafraction motion using a single kV imager during radiotherapy is feasible and can be performed with acceptable accuracy.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app