Comparison of kilovoltage cone-beam computed tomography with megavoltage projection pairs for paraspinal radiosurgery patient alignment and position verification

Sergey A Kriminski, D Michael Lovelock, Venkatraman E Seshan, Imad Ali, Peter Munro, Howard I Amols, Zvi Fuks, Mark Bilsky, Yoshiya Yamada
International Journal of Radiation Oncology, Biology, Physics 2008 August 1, 71 (5): 1572-80
18539405

PURPOSE: Implanted gold markers and megavoltage (MV) portal imaging are commonly used for setup verification of paraspinal tumors treated with high-dose, single-fraction radiotherapy. We investigated whether the use of kilovoltage cone-beam computed tomography (CBCT) imaging eliminates the need for marker implantation.

METHODS AND MATERIALS: Patients with paraspinal disease who were eligible for single-fraction stereotactic body radiotherapy were accrued to an institutional review board-approved protocol. Each of 16 patients underwent implantation of fiducial markers near the target. The markers were visible on the MV images. Three MV image pairs were acquired for each patient (initial, verification, and final) and were registered to the reference images. Every MV pair was complemented by a CBCT scan. CBCT image registration was performed automatically by maximizing the mutual information using a region of interest that excluded the markers. The corrections, as determined from the MV images, were compared with these from CBCT and were used for actual patient setup.

RESULTS: The mean and standard deviation of the absolute values of the differences between the CBCT and MV corrections were 1.0 +/- 0.7, 1.0 +/- 0.6, and 1.0 +/- 0.8 mm for the left-right, anteroposterior, and superoinferior directions, respectively. The absolute differences between the corresponding pre- and post-treatment kilovoltage CBCT image registration were 0.6 +/- 0.5, 0.6 +/- 0.5, and 1.0 +/- 0.8 mm.

CONCLUSION: The setup corrections found using CBCT without the use of implanted markers were consistent with the marker registration on MV projections. CBCT has additional advantages, including better positioning precision and robust automatic three-dimensional registration, as well as eliminating the need for invasive marker implantation. We have adopted CBCT for the setup of all single-fraction paraspinal patients. Our data have also demonstrated that target displacements during treatment are insignificant.

Comments

You need to log in or sign up for an account to be able to comment.

No comments yet, be the first to post one!

Related Papers

Available on the App Store

Available on the Play Store
Remove bar
Read by QxMD icon Read
18539405
×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"