Estimation of setup uncertainty using planar and MVCT imaging for gynecologic malignancies

Lakshmi Santanam, Jacqueline Esthappan, Sasa Mutic, Eric E Klein, S Murty Goddu, Summer Chaudhari, Sasha Wahab, Issam M El Naqa, Daniel A Low, Perry W Grigsby
International Journal of Radiation Oncology, Biology, Physics 2008 August 1, 71 (5): 1511-7
18538499

PURPOSE: This prospective study investigates gynecologic malignancy online treatment setup error corrections using planar kilovoltage/megavoltage (KV/MV) imaging and helical MV computed tomography (MVCT) imaging.

METHODS AND MATERIALS: Twenty patients were divided into two groups. The first group (10 patients) was imaged and treated using a conventional linear accelerator (LINAC) with image-guidance capabilities, whereas the second group (10 patients) was treated using tomotherapy with MVCT capabilities. Patients treated on the LINAC underwent planar KV and portal MV imaging and a two-dimensional image registration algorithm was used to match these images to digitally reconstructed radiographs (DRRs). Patients that were treated using tomotherapy underwent MVCT imaging, and a three-dimensional image registration algorithm was used to match planning CT to MVCT images. Subsequent repositioning shifts were applied before each treatment and recorded for further analysis. To assess intrafraction motion, 5 of the 10 patients treated on the LINAC underwent posttreatment planar imaging and DRR matching. Based on these data, patient position uncertainties along with estimated margins based on well-known recipes were determined.

RESULTS: The errors associated with patient positioning ranged from 0.13 cm to 0.38 cm, for patients imaged on LINAC and 0.13 cm to 0.48 cm for patients imaged on tomotherapy. Our institutional clinical target volume-PTV margin value of 0.7 cm lies inside the confidence interval of the margins established using both planar and MVCT imaging.

CONCLUSION: Use of high-quality daily planar imaging, volumetric MVCT imaging, and setup corrections yields excellent setup accuracy and can help reduce margins for the external beam treatment of gynecologic malignancies.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
18538499
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"