We have located links that may give you full text access.
COMPARATIVE STUDY
JOURNAL ARTICLE
Measurement of intraocular pressure in LASIK and LASEK patients using the Reichert Ocular Response Analyzer and Goldmann applanation tonometry.
Journal of Refractive Surgery 2008 April
PURPOSE: To determine the efficacy of the Reichert Ocular Response Analyzer (ORA) to measure intraocular pressure (IOP) following corneal laser refractive surgery.
METHODS: Intraocular pressure was measured using Goldmann applanation tonometry preoperatively and 3 months following LASIK and LASEK for all levels of myopia and low levels of hyperopia. In LASIK eyes, 120-microm flaps were cut using the Hansatome XP. The ORA was used to measure Goldmann correlated IOP and corneal compensated IOP 3 months postoperatively. Postoperative central corneal thickness, hysteresis, and corneal curvature were assessed to determine their relationship with postoperative change in Goldmann applanation tonometry IOP. Efficacy of ORA IOP measurement was analyzed by comparing Goldmann applanation tonometry IOP with Goldmann correlated IOP.
RESULTS: LASEK was performed on 35 eyes, and LASIK was performed on 90 eyes. In the LASIK group, mean Goldmann applanation tonometry IOP decreased 3.7+/-2.3 mmHg postoperatively (P=.00). Postoperative Goldmann applanation tonometry did not differ significantly (P=.06) from postoperative ORA Goldmann correlated IOP (10.2+/-2.1 mmHg). In the LASEK group, mean Goldmann applanation tonometry IOP decreased 3.9+/-2.3 mmHg (P=.00). Postoperative Goldmann applanation tonometry did not differ significantly (P=.6) from postoperative ORA Goldmann correlated IOP (10.7+/-2.5 mmHg). Postoperative decrease in Goldmann applanation tonometry IOP did not correlate with age, ablation depth, pre- and postoperative central corneal thickness or corneal hysteresis, or postoperative decrease in corneal curvature.
CONCLUSIONS: Intraocular pressure measurements were similar using both the ORA Goldmann correlated IOP and Goldmann applanation tonometry following keratorefractive surgery.
METHODS: Intraocular pressure was measured using Goldmann applanation tonometry preoperatively and 3 months following LASIK and LASEK for all levels of myopia and low levels of hyperopia. In LASIK eyes, 120-microm flaps were cut using the Hansatome XP. The ORA was used to measure Goldmann correlated IOP and corneal compensated IOP 3 months postoperatively. Postoperative central corneal thickness, hysteresis, and corneal curvature were assessed to determine their relationship with postoperative change in Goldmann applanation tonometry IOP. Efficacy of ORA IOP measurement was analyzed by comparing Goldmann applanation tonometry IOP with Goldmann correlated IOP.
RESULTS: LASEK was performed on 35 eyes, and LASIK was performed on 90 eyes. In the LASIK group, mean Goldmann applanation tonometry IOP decreased 3.7+/-2.3 mmHg postoperatively (P=.00). Postoperative Goldmann applanation tonometry did not differ significantly (P=.06) from postoperative ORA Goldmann correlated IOP (10.2+/-2.1 mmHg). In the LASEK group, mean Goldmann applanation tonometry IOP decreased 3.9+/-2.3 mmHg (P=.00). Postoperative Goldmann applanation tonometry did not differ significantly (P=.6) from postoperative ORA Goldmann correlated IOP (10.7+/-2.5 mmHg). Postoperative decrease in Goldmann applanation tonometry IOP did not correlate with age, ablation depth, pre- and postoperative central corneal thickness or corneal hysteresis, or postoperative decrease in corneal curvature.
CONCLUSIONS: Intraocular pressure measurements were similar using both the ORA Goldmann correlated IOP and Goldmann applanation tonometry following keratorefractive surgery.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app