We have located links that may give you full text access.
COMPARATIVE STUDY
EVALUATION STUDIES
JOURNAL ARTICLE
Cardiac output measurement in patients undergoing liver transplantation: pulmonary artery catheter versus uncalibrated arterial pressure waveform analysis.
Anesthesia and Analgesia 2008 May
BACKGROUND: Cardiac output (CO) and invasive hemodynamic measurements are useful during liver transplantation. The pulmonary artery catheter (PAC) is commonly used for these patients, despite the potential complications. Recently, a less invasive device (Vigileo/FloTrac) became available, which estimates CO using arterial pressure waveform analysis without external calibration. In this study, we compared CO obtained with a PAC using automatic thermodilution, instantaneous CO stat-mode (ICO(SM)), and CO obtained with the new device, arterial pressure waveform analysis (APCO) in patients undergoing liver transplantation.
METHODS: Twenty sets of simultaneous measurements of APCO and ICO(SM) were determined in sedated and mechanically ventilated patients undergoing liver transplantation. Time points were as follows: after PAC insertion (T1-3), after portal clamping (T4-6), during anhepathy (T7-9), after graft reperfusion (T10-15), and in the postoperative period in the intensive care unit (T15-20).
RESULTS: We enrolled 20 patients and 400 measurements were obtained. No data were rejected. Bias between ICO(SM) and APCO was 0.8 L/min, 95% limits of agreement were -1.8 to 3.5 L/min. The percentage error was 43%. Bias between ICO(SM) and APCO was correlated with systemic vascular resistance [r(2) = 0.55, P < 0.0001, y = 15.8-2.2 ln(x)] and subgroup analysis revealed an increase in the bias and in the percentage error in patients with low systemic vascular resistance (Child-Pugh grade B and C patients). There was no difference between the different surgical periods.
CONCLUSIONS: Our results suggest that Vigileo/FloTrac CO monitoring data do not agree well with those of automatic thermodilution in patients undergoing liver transplantation, especially in Child-Pugh grade B and C patients with low systemic vascular resistance.
METHODS: Twenty sets of simultaneous measurements of APCO and ICO(SM) were determined in sedated and mechanically ventilated patients undergoing liver transplantation. Time points were as follows: after PAC insertion (T1-3), after portal clamping (T4-6), during anhepathy (T7-9), after graft reperfusion (T10-15), and in the postoperative period in the intensive care unit (T15-20).
RESULTS: We enrolled 20 patients and 400 measurements were obtained. No data were rejected. Bias between ICO(SM) and APCO was 0.8 L/min, 95% limits of agreement were -1.8 to 3.5 L/min. The percentage error was 43%. Bias between ICO(SM) and APCO was correlated with systemic vascular resistance [r(2) = 0.55, P < 0.0001, y = 15.8-2.2 ln(x)] and subgroup analysis revealed an increase in the bias and in the percentage error in patients with low systemic vascular resistance (Child-Pugh grade B and C patients). There was no difference between the different surgical periods.
CONCLUSIONS: Our results suggest that Vigileo/FloTrac CO monitoring data do not agree well with those of automatic thermodilution in patients undergoing liver transplantation, especially in Child-Pugh grade B and C patients with low systemic vascular resistance.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app