Rebound tonometry in conscious, conditioned mice avoids the acute and profound effects of anesthesia on intraocular pressure.

AIMS: The aims of this study were to evaluate the accuracy, repeatability, and safety of multiple intraocular pressure (IOP) measurements by a commercially available rebound tonometer in conscious, conditioned mice, and to characterize the acute and profound effects of anesthesia on IOP in mice.

METHODS: To test the accuracy of the tonometer, IOPs of CD-1 mice under ketamine/xylazine anesthesia were experimentally set and monitored with a water manometer/transducer system following transcorneal cannulation while simultaneously performing tonometry. The long- and short-term repeatability of the tonometer was tested in conscious, restrained mice, as measurements were taken once-daily in the afternoon for 4 consecutive days. On day 5, IOPs were measured in the same mice once every 4 min for 32 min. On 2 separate days, mice were administered ketamine/xylazine or 2,2,2-tribromoethanol anesthesia, in a crossover design, and IOPs were measured once every 2 min for 32 min. Rebound tonometry was performed in conscious mice before and 1 hour after 1 drop of timolol maleate (10 microL of 0.5%) application to 1 eye.

RESULTS: IOP measurements by rebound tonometry correlated well with manometry for pressures between 8 and 38 mmHg (y = 0.98x - 0.32, R(2) = 0.94; P < 0.001). The average tonometric IOP was invariant over 4 days (range, 11.7-13.2 mmHg). IOPs dropped significantly ( P < or = 0.05) within 6 min (ketamine/xylazine) or 10 min (2,2,2-tribromoethanol) postadministration of anesthesia but not with conscious restraint. Timolol significantly (P < 0.001) lowered IOP from 12.8 +/- 0.3 (mean +/- standard error of the mean) to 10.1 +/- 0.6 mmHg, as measured by the tonometer.

CONCLUSIONS: Rebound tonometry can be used to obtain accurate IOP measurements in conscious, restrained mice while avoiding the rapid and profound ocular hypotensive effects of general anesthesia. Small changes in IOP with an aqueous-flow suppressant are readily detectable with conscious restraint that may be missed with chemical restraint.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app