COMPARATIVE STUDY
JOURNAL ARTICLE
Transoesophageal echo-Doppler vs. thermodilution cardiac output measurement during hepatic vascular exclusion in liver transplantation.
European Journal of Anaesthesiology 2008 June
BACKGROUND AND OBJECTIVE: Continuous monitoring of cardiac output during liver transplantation is essential to evaluate the patient's haemodynamic tolerance to acute volume variations. The aim of this study was to compare the cardiac output values obtained with a transoesophageal echo-Doppler and those obtained with a continuous thermodilution cardiac output pulmonary artery catheter.
METHODS: Twenty adult patients were prospectively studied during a 5 min hepatic vascular exclusion test performed at the end of the dissection phase. Echo-Doppler and continuous thermodilution cardiac output, mean arterial pressure and end-tidal CO2 were measured before and at the end of the test.
RESULTS: Before the test, echo-Doppler cardiac output was 7.0 +/- 2.7 L min(-1) and thermodilution was 9.4 +/- 3.1 L min(-1), (R = 0.85, P < 0.001). The end test values were, respectively, 3.5 +/- 2.7 and 7.8 +/- 3.5 L min(-1) (R = 0.23, P = 0.34). Bland and Altman analysis showed a bias of -2.2 before the test, which increased to -4.4 at the end of the test. Mean arterial pressure decreased from 85.5 +/- 15 to 66.8 +/- 16 mmHg, end-tidal CO2 from 31.4 +/- 2.3 to 23.8 +/- 2.7 mmHg.
CONCLUSION: Echo-Doppler cardiac output values are different from those measured by thermodilution cardiac output in these patients. Echo-Doppler cardiac output monitoring seems to detect the output changes, which can occur during acute haemodynamic changes more rapidly than thermodilution cardiac output in the course of liver transplantation.
METHODS: Twenty adult patients were prospectively studied during a 5 min hepatic vascular exclusion test performed at the end of the dissection phase. Echo-Doppler and continuous thermodilution cardiac output, mean arterial pressure and end-tidal CO2 were measured before and at the end of the test.
RESULTS: Before the test, echo-Doppler cardiac output was 7.0 +/- 2.7 L min(-1) and thermodilution was 9.4 +/- 3.1 L min(-1), (R = 0.85, P < 0.001). The end test values were, respectively, 3.5 +/- 2.7 and 7.8 +/- 3.5 L min(-1) (R = 0.23, P = 0.34). Bland and Altman analysis showed a bias of -2.2 before the test, which increased to -4.4 at the end of the test. Mean arterial pressure decreased from 85.5 +/- 15 to 66.8 +/- 16 mmHg, end-tidal CO2 from 31.4 +/- 2.3 to 23.8 +/- 2.7 mmHg.
CONCLUSION: Echo-Doppler cardiac output values are different from those measured by thermodilution cardiac output in these patients. Echo-Doppler cardiac output monitoring seems to detect the output changes, which can occur during acute haemodynamic changes more rapidly than thermodilution cardiac output in the course of liver transplantation.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app