Percutaneous treatment for mitral regurgitation: the QuantumCor system.

AIMS: Percutaneous edge-to-edge techniques and annuloplasty have been used to treat mitral regurgitation (MR). However, neither intervention can be performed reliably a second time and, with annuloplasty, a foreign body is left behind. The mitral and tricuspid annuli are areas of dense collagen (Fig. 1); treatment with radiofrequency (RF) energy in sheep reduces their size, and can be repeated without affecting the coronary sinus. RF energy may also be used in leaflet procedures. Our aim was to improve mitral valve competence using techniques that can be incorporated into a minimally invasive approach.

METHODS: In open-heart procedures in 16 healthy sheep (6 with naturally occurring MR), we used a malleable probe (QuantumCor, Inc., Lake Forest, CA) that conforms to the annular shape to deliver RF energy via a standard generator to replicate a surgical mitral annular ring. Four segments of the posterior mitral valve annulus were treated while on cardiopulmonary support via a left thoracotomy with access via the atrial appendage. Seven sheep were followed chronically.

RESULTS: All sheep underwent intracardiac echocardiography (ICE) or direct circumferential measurement of the mitral annulus before and after RF therapy. RF therapy was administered in less than 4 minutes in each case, and the mean anteroposterior (AP) annular distance was reduced by a mean of 5.75 +/- 0.86 mm (23.8% reduction, P< 0.001). In the 6 sheep with nonischemic MR, regurgitation was eliminated. Acute histopathology (HP) demonstrated no damage to the leaflets, coronary sinuses, or coronary arteries. At 30 days, the AP distance continued to be reduced in the 7 surviving sheep (mean 5.0 +/- .6 mm, 21.4% reduction, P< 0.001).

CONCLUSIONS: In a sheep model, RF energy applied for less than 4 minutes per case at subablative temperatures in four quadrants of the posterior mitral valve annulus reduced the AP and circumferential annular distances significantly, and eliminated nonischemic MR. Results will need to be confirmed in follow-up studies to determine safety and efficacy. RF energy administered as a novel, percutaneous method of mitral valve annuloplasty may have the potential to reduce morbidity and mortality associated with current surgical techniques.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app