Efficacy of adaptive servoventilation in treatment of complex and central sleep apnea syndromes.
Chest 2007 December
BACKGROUND: Complex sleep apnea syndrome (CompSAS) is recognized by the concurrence of mixed or obstructive events with central apneas, the latter predominating on exposure to continuous positive airway pressure (CPAP). Treatment of CompSAS or central sleep apnea (CSA) syndrome with adaptive servoventilation (ASV) is now an option, but no large series exist describing the application and effectiveness of ASV.
METHODS: Retrospective chart review of the first 100 patients who underwent polysomnography using ASV at Mayo Clinic Sleep Center.
RESULTS: ASV titration was performed for CompSAS (63%), CSA (22%), or CSA/Cheyne Stokes breathing patterns (15%). The median diagnostic sleep apnea hypopnea index (AHI) was 48 events per hour (range, 24 to 62). With CPAP, obstructive apneas decreased, but the appearance of central apneas maintained the AHI at 31 events per hour (range, 17 to 47) [p = 0.02]. With bilevel positive airway pressure (BPAP) in spontaneous mode, AHI trended toward worsening vs baseline, with a median of 75 events per hour (range, 46 to 111) [p = 0.055]. BPAP with a backup rate improved the AHI to 15 events per hour (range, 11 to 31) [p = 0.002]. Use of ASV dramatically improved the AHI to a mean of 5 events per hour (range, 1 to 11) vs baseline and vs CPAP (p < 0.0001). ASV also resulted in an increase in rapid eye movement sleep vs baseline and CPAP (18% vs 12% and 10%, respectively; p < 0.0001). Overall, 64 patients responded to the ASV treatment with a mean AHI < 10 events per hour. Of the 44 successful survey follow-up patients contacted, 32 patients reported some improvement in sleep quality.
CONCLUSION: The ASV device appears to be an effective treatment of both CompSAS and CSA syndromes that are resistant to CPAP.
METHODS: Retrospective chart review of the first 100 patients who underwent polysomnography using ASV at Mayo Clinic Sleep Center.
RESULTS: ASV titration was performed for CompSAS (63%), CSA (22%), or CSA/Cheyne Stokes breathing patterns (15%). The median diagnostic sleep apnea hypopnea index (AHI) was 48 events per hour (range, 24 to 62). With CPAP, obstructive apneas decreased, but the appearance of central apneas maintained the AHI at 31 events per hour (range, 17 to 47) [p = 0.02]. With bilevel positive airway pressure (BPAP) in spontaneous mode, AHI trended toward worsening vs baseline, with a median of 75 events per hour (range, 46 to 111) [p = 0.055]. BPAP with a backup rate improved the AHI to 15 events per hour (range, 11 to 31) [p = 0.002]. Use of ASV dramatically improved the AHI to a mean of 5 events per hour (range, 1 to 11) vs baseline and vs CPAP (p < 0.0001). ASV also resulted in an increase in rapid eye movement sleep vs baseline and CPAP (18% vs 12% and 10%, respectively; p < 0.0001). Overall, 64 patients responded to the ASV treatment with a mean AHI < 10 events per hour. Of the 44 successful survey follow-up patients contacted, 32 patients reported some improvement in sleep quality.
CONCLUSION: The ASV device appears to be an effective treatment of both CompSAS and CSA syndromes that are resistant to CPAP.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app