Accuracy of cardiac output measurements with pulse contour analysis (PulseCO) and Doppler echocardiography during off-pump coronary artery bypass grafting

C Missant, S Rex, P F Wouters
European Journal of Anaesthesiology 2008, 25 (3): 243-8
17996125

BACKGROUND AND OBJECTIVE: During off-pump coronary bypass grafting, surgical manipulation and dislocation of the heart may cause cardiovascular instability. Monitoring of cardiac output facilitates intraoperative haemodynamic management but pulmonary artery catheters are often considered too invasive. Pulse contour analysis and transoesophageal echocardiography could serve as alternatives, but there is controversy about their accuracies. We validated pulse contour analysis using a standard radial arterial catheter (PulseCO) and aortic Doppler flowmetry with transoesophageal echocardiography in patients undergoing off-pump coronary bypass surgery. Pulmonary arterial thermodilution served as the reference technique.

METHODS: In 20 patients undergoing off-pump coronary bypass, cardiac output was measured with bolus thermodilution (COTD), pulse contour analysis (COPC), and transoesophageal echocardiography (COecho) at fixed time intervals during the procedure. Data were compared using linear regression and Bland-Altman analysis. At the end of the procedure, dobutamine was infused at a rate of 2.5 microg kg(-1) min(-1) in six patients to study the agreement between methods in quantifying changes in cardiac output.

RESULTS: Comparison between COPC and COTD showed a bias +/- limits of agreement of -0.03 +/- 1.30 L min(-1) (mean error 29%). Doppler echocardiography was not always feasible when the heart was displaced from the oesophagus and had lower accuracy: bias +/- limits of agreement vs. COTD was 0.45 +/- 1.93 (mean error 43%). Increases in cardiac output induced by dobutamine were well quantified both by pulse contour analysis (COPC = 0.76 x COTD + 0.58; r(2) = 0.65) and Doppler, although the latter tended to overestimate these changes (COecho = 1.58 x COTD - 0.13; r(2) = 0.53).

CONCLUSIONS: Calibrated pulse contour analysis using the PulseCO system is an acceptable technique to measure cardiac output non-invasively in off-pump coronary bypass patients. Doppler echocardiography performs less well and is not always feasible with transoesophageal echocardiography when the heart is displaced.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Trending Papers

Remove bar
Read by QxMD icon Read
17996125
×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"