Image quality and radiation exposure at pulmonary CT angiography with 100- or 120-kVp protocol: prospective randomized study.

Radiology 2007 November
PURPOSE: To prospectively compare 16-section multidetector computed tomography (CT) at 100 and 120 kVp for image quality and radiation dose.

MATERIALS AND METHODS: The study had institutional review board approval; written informed consent was obtained. Sixty patients were referred for evaluation of suspected pulmonary embolism with CT angiography. Patients were randomly assigned to a 100-kVp (n = 30; 17 men, 13 women; mean age, 66 years +/- 17 [standard deviation]; range, 19-89 years) or 120-kVp (n = 30; 15 men, 15 women; mean age, 62 years +/- 15; range, 28-86 years) protocol. Other scanning parameters were kept constant. Contrast medium was injected automatically with bolus tracking. Pulmonary vessel enhancement and image noise were quantified; signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated. Subjective vessel contrast was assessed by two radiologists in consensus. Effective dose was calculated on the basis of dose length product and volume CT dose index. Results of both protocols were compared by using the chi(2) test and Student t test.

RESULTS: The 100-kVp protocol had a nonsignificantly higher mean vessel attenuation than the 120-kVp protocol (386.8 HU +/- 130.1 vs 317.9 HU +/- 112.5; P = .56) and a nonsignificantly higher image noise (16.9 HU +/- 5.8 vs 13.7 HU +/- 6.2; P = .84), which resulted in almost identical SNR (25.3 +/- 11.7 vs 27.0 +/- 14.5; P = .37) and CNR (22.0 +/- 11.2 vs 22.9 +/- 13.1; P = .51). There was no significant difference in subjective image quality between protocols. Mean effective dose for the 100-kVp protocol was significantly lower than that for the 120-kVp protocol (1.37 mSv +/- 0.39 vs 2.44 mSv +/- 0.97; -44%; P < .001).

CONCLUSION: Reduction of kilovoltage from 120 to 100 kVp resulted in significant reduction of effective dose at pulmonary CT angiography, without significant loss of objective or subjective image quality.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app