Add to saved papers

Accuracy of cerebral monitoring in detecting cerebral ischemia during carotid endarterectomy: a comparison of transcranial Doppler sonography, near-infrared spectroscopy, stump pressure, and somatosensory evoked potentials.

Anesthesiology 2007 October
BACKGROUND: This study compares the accuracy of cerebral monitoring systems in detecting cerebral ischemia during carotid endarterectomy.

METHODS: The authors compared transcranial Doppler sonography (TCD), near-infrared spectroscopy (NIRS), stump pressure (SP) measurement, and somatosensory evoked potentials (SEP) in 48 patients undergoing carotid surgery during regional anesthesia. Cerebral ischemia was assumed when neurologic deterioration occurred. During clamping, the minimum mean middle cerebral artery velocity (TCD(min)), its percentage change (TCD%), the minimum regional saturation of oxygen (NIRS(min)), its percentage change (NIRS%), the mean SP, and the changes of SEP amplitude were recorded. To analyze the corresponding sensitivity and specificity of each parameter, the authors performed receiver operating characteristic analysis.

RESULTS: Neurologic deterioration occurred in 12 patients. SP and NIRS were successfully performed in all patients. TCD monitoring was not possible in 10 (21%); SEP was not possible in 2 patients (4%). All parameters provided the ability to distinguish between ischemic and nonischemic patients. TCD% and NIRS% showed significantly better discrimination than TCD(min) and NIRS(min) (P < 0.05). The highest area under the curve (AUC) was found for TCD% (AUC = 0.973), but there was no significant difference compared with NIRS% (AUC = 0.905) and SP (AUC = 0.925). The lowest AUC was found for SEP (AUC = 0.749), which was significantly lower than that for TCD%, NIRS%, and SP.

CONCLUSIONS: TCD%, NIRS%, and SP measurement provide similar accuracy for the detection of cerebral ischemia during carotid surgery. Lower accuracy was found for SEP monitoring. Because of the high rate of technical difficulties (21%), TCD monitoring was the least practical of the investigated monitoring devices.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app