Assessment of the residual error in soft tissue setup in patients undergoing partial breast irradiation: results of a prospective study using cone-beam computed tomography

Toyosi Fatunase, Zhiheng Wang, Sua Yoo, Jessica L Hubbs, Robert G Prosnitz, Fang-Fang Yin, Lawrence B Marks
International Journal of Radiation Oncology, Biology, Physics 2008 March 15, 70 (4): 1025-34
17892919

PURPOSE: On-board cone-beam computed tomography (CBCT) provides soft tissue information that may improve setup accuracy in patients undergoing accelerated partial breast irradiation (APBI). We used CBCT to assess the residual error in soft tissue after two-dimensional kV/MV alignment based on bony anatomy. We also assessed the dosimetric impact of this error.

METHODS AND MATERIALS: Ten patients undergoing APBI were studied as part of an institutional review board-approved prospective trial. Patients were aligned based on skin/cradle marks plus orthogonal kV/MV images registered based on bony landmarks to digitally reconstructed radiographs from the planning CT. A subsequent CBCT was registered to the planning CT using soft tissue information. This "residual error" and its dosimetric impact was measured.

RESULTS: The root-mean-square of the residual error was 3, 4, and 4 mm, in the right-left, anterior-posterior, and superior-inferior directions, respectively. The average vector sum was 6+/-2 mm. Average reductions in mean dose to the lumpectomy cavity, clinical target volume (CTV), and planning target volume were 0.1%, 0.4%, and 1%, respectively. The mean difference in the clinical target and planning target volumes that received 95% of the prescribed dose (V95) were 1% and 4%.

CONCLUSIONS: In this initial study with a modest number of patients, the residual error in soft tissue was typically <5 mm, and with the field margins used, the resultant dosimetric consequences were modest. In patients immobilized in a customized cradle, setup using orthogonal kV images thus appears accurate and reproducible. The CBCT technique may have particular utility in patients with larger breast volumes or breast deformations. Further studies involving larger numbers of patients are needed to further assess the utility of CBCT.

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Available on the App Store

Available on the Play Store
Remove bar
Read by QxMD icon Read
17892919
×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"