Clinical Trial
Comparative Study
Journal Article
Multicenter Study
Validation Studies
Add like
Add dislike
Add to saved papers

Validation of a continuous, arterial pressure-based cardiac output measurement: a multicenter, prospective clinical trial.

INTRODUCTION: The present study compared measurements of cardiac output by an arterial pressure-based cardiac output (APCO) analysis method with measurement by intermittent thermodilution cardiac output (ICO) via pulmonary artery catheter in a clinical setting.

METHODS: The multicenter, prospective clinical investigation enrolled patients with a clinical indication for cardiac output monitoring requiring pulmonary artery and radial artery catheters at two hospitals in the United States, one hospital in France, and one hospital in Belgium. In 84 patients (69 surgical patients), the cardiac output was measured by analysis of the arterial pulse using APCO and was measured via pulmonary artery catheter by ICO; to establish a reference comparison, the cardiac output was measured by continuous cardiac output (CCO). Data were collected continuously by the APCO and CCO technologies, and at least every 4 hours by ICO. No clinical interventions were made as part of the study.

RESULTS: For APCO compared with ICO, the bias was 0.20 l/min, the precision was +/- 1.28 l/min, and the limits of agreement were -2.36 l/m to 2.75 l/m. For CCO compared with ICO, the bias was 0.66 l/min, the precision was +/- 1.05 l/min, and the limits of agreement were -1.43 l/m to 2.76 l/m. The ability of APCO and CCO to assess changes in cardiac output was compared with that of ICO. In 96% of comparisons, APCO tracked the change in cardiac output in the same direction as ICO. The magnitude of change was comparable 59% of the time. For CCO, 95% of comparisons were in the same direction, with 58% of those changes being of similar magnitude.

CONCLUSION: In critically ill patients in the intensive care unit, continuous measurement of cardiac output using either APCO or CCO is comparable with ICO. Further study in more homogeneous populations may refine specific situations where APCO reliability is strongest.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app