Balanced steady-state free precession vs. segmented fast low-angle shot for the evaluation of ventricular volumes, mass, and function at 3 Tesla.
Journal of Magnetic Resonance Imaging : JMRI 2007 August
PURPOSE: To compare balanced steady-state free precession (SSFP) and segmented fast low angle shot (FLASH) for quantification of left and right ventricular volumes and function and for left ventricular mass at high field (3 Tesla).
MATERIALS AND METHODS: A total of 33 patients (19 male, mean age 54 years) with various forms of heart disease underwent ventricular function studies using cine SSFP and FLASH sequences with identical slice orientations.
RESULTS: Using SSFP, left ventricular end-diastolic (+10 mL [4.7%], P < 0.001) and end-systolic volumes (+9 mL [6.1%], P < 0.001) measured larger whereas mass was considerably smaller (-23 g [-12.9%], P < 0.001) and ejection fraction (-1% [-3.2%], P < 0.01) marginally smaller. Right ventricular end-diastolic (+4 mL [2.6%], P = 0.001) and end-systolic volumes (+4 mL [5.1%], P < 0.01) were also larger, but no significant difference for right ventricular ejection fraction (P = 0.05) was found.
CONCLUSION: Similar to previous results at 1.5 Tesla, at high magnetic field the cine SSFP technique led to discrete but significantly higher ventricular volume measurements and to a significantly smaller measurement of left ventricular mass in patients. The effect on left and right ventricular ejection fraction was minor, although the difference remained significant for the left ventricle.
MATERIALS AND METHODS: A total of 33 patients (19 male, mean age 54 years) with various forms of heart disease underwent ventricular function studies using cine SSFP and FLASH sequences with identical slice orientations.
RESULTS: Using SSFP, left ventricular end-diastolic (+10 mL [4.7%], P < 0.001) and end-systolic volumes (+9 mL [6.1%], P < 0.001) measured larger whereas mass was considerably smaller (-23 g [-12.9%], P < 0.001) and ejection fraction (-1% [-3.2%], P < 0.01) marginally smaller. Right ventricular end-diastolic (+4 mL [2.6%], P = 0.001) and end-systolic volumes (+4 mL [5.1%], P < 0.01) were also larger, but no significant difference for right ventricular ejection fraction (P = 0.05) was found.
CONCLUSION: Similar to previous results at 1.5 Tesla, at high magnetic field the cine SSFP technique led to discrete but significantly higher ventricular volume measurements and to a significantly smaller measurement of left ventricular mass in patients. The effect on left and right ventricular ejection fraction was minor, although the difference remained significant for the left ventricle.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app