COMPARATIVE STUDY
JOURNAL ARTICLE

Reconstructed versus real-time 3-dimensional echocardiography: comparison with magnetic resonance imaging

Carly Jenkins, Rodel Leano, Jonathan Chan, Thomas H Marwick
Journal of the American Society of Echocardiography 2007, 20 (7): 862-8
17617313

BACKGROUND: The accuracy, reproducibility, and test-retest reliability of 3-dimensional (3D) echocardiography (3DE) with 3D reconstruction (3DR) and real-time (RT) imaging (RT 3DE) exceed that of 2-dimensional echocardiography (2DE). However, image quality with RT 3DE is inferior to 2DE and we sought to determine whether this justified ongoing use of 3DR.

METHODS: Unselected patients (n = 30, 22 men, age 66 +/- 7 years) presenting to the echocardiography laboratory for left ventricular (LV) evaluation were studied with 2DE and RT 3DE; 3DR images were obtained using external localization. The 3D measurements and reconstructions were obtained offline. Magnetic resonance images (MRI) were obtained using true free induction, steady state precession during breath hold and 3D volumes and ejection fraction (EF) were measured using 3D software. A separate cohort of 20 patients (13 men, age 60 +/- 12 years) was measured for test-retest variation.

RESULTS: All echocardiographic measures underestimated LV volumes and EF compared with MRI, but this was least with RT 3DE. End-diastolic volume by MRI (168 +/- 54 mL) was underestimated by RT 3DE (-15 +/- 31, P = .02), 3DR (-26 +/- 33, P < .01), and 2DE (-57 +/- 40, P < .01). Similarly, end-systolic volume by MRI (86 +/- 50 mL) was underestimated by RT 3DE (-15 +/- 31, P = .02), 3DR (-26 +/- 33, P < .01), and 2DE (-57 +/- 40, P < .01). However, EF measurements were similar with each method. Test-retest variation was less and interobserver and intraobserver correlations were better with RT 3DE for volumes and EF, compared with 3DR and 2DE.

CONCLUSIONS: Despite limitations of image quality, RT 3DE is the most feasible and accurate approach for LV volume and EF measurements and follow-up LV assessment in daily practice.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
17617313
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"