We have located links that may give you full text access.
Evaluation Studies
Journal Article
A prospective clinical investigation of the failure rate of immediately loaded mini-implants used for orthodontic anchorage.
AIM: To evaluate the failure rate of immediately loaded mini-implants used for orthodontic anchorage.
MATERIALS AND METHODS: 140 mini-implants were inserted in 98 patients and immediately loaded with 50cN super-elastic coil springs. 99 devices were inserted in the lower jaw and 41 in the upper jaw in the following sites: mandibular symphusis (n=17), mandibular alveolar process (n=72), maxillary alveolar process (n=32), retromolar area (n=10), and palate (n=9). The mini-implants were used to perform the following dental movements: molar uprighting, molar uprighting and mesial movement, molar mesial movement, incisor intrusion and proclination, incisor retraction, premolar intrusion, midline correction, premolar distal movement, and molar intrusion. All devices that showed complete absence of mobility after 120 days of continuous load were scored as successful. Those that showed minimum mobility but stayed in place and could resist further load were scored as partial failures, while those that were lost were scored as failures.
RESULTS: 13 of the mini-implants failed (9.3%), 9 were partial failures (6.4%). The upper jaw had a greater failure rate (12.2%) than the lower jaw (8.0%). The palate was the anatomic location with the greatest risk of failure, while in the mandibular alveolar process the lowest failure rate was observed.
CONCLUSIONS: The overall failure rate was similar to other investigations where a healing period had been performed, suggesting that immediate loading with light forces should not be considered a risk factor. Other factors such as inflammation of the surrounding soft tissues, bone characteristics, thickness of the mucosa and incorrect surgical procedure should be considered determinants of clinical failure.
MATERIALS AND METHODS: 140 mini-implants were inserted in 98 patients and immediately loaded with 50cN super-elastic coil springs. 99 devices were inserted in the lower jaw and 41 in the upper jaw in the following sites: mandibular symphusis (n=17), mandibular alveolar process (n=72), maxillary alveolar process (n=32), retromolar area (n=10), and palate (n=9). The mini-implants were used to perform the following dental movements: molar uprighting, molar uprighting and mesial movement, molar mesial movement, incisor intrusion and proclination, incisor retraction, premolar intrusion, midline correction, premolar distal movement, and molar intrusion. All devices that showed complete absence of mobility after 120 days of continuous load were scored as successful. Those that showed minimum mobility but stayed in place and could resist further load were scored as partial failures, while those that were lost were scored as failures.
RESULTS: 13 of the mini-implants failed (9.3%), 9 were partial failures (6.4%). The upper jaw had a greater failure rate (12.2%) than the lower jaw (8.0%). The palate was the anatomic location with the greatest risk of failure, while in the mandibular alveolar process the lowest failure rate was observed.
CONCLUSIONS: The overall failure rate was similar to other investigations where a healing period had been performed, suggesting that immediate loading with light forces should not be considered a risk factor. Other factors such as inflammation of the surrounding soft tissues, bone characteristics, thickness of the mucosa and incorrect surgical procedure should be considered determinants of clinical failure.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Essential thrombocythaemia: A contemporary approach with new drugs on the horizon.British Journal of Haematology 2024 April 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app