Errors in the computerized electrocardiogram interpretation of cardiac rhythm

Atman P Shah, Stanley A Rubin
Journal of Electrocardiology 2007, 40 (5): 385-90
17531257

BACKGROUND: More than 100 million computer-interpreted electrocardiograms (ECG-C) are obtained annually. However, there are few contemporary published data on the accuracy of cardiac rhythm interpretation by this method.

PURPOSE: The purpose of this study is to determine the accuracy of ECG-C rhythm interpretation in a typical patient population.

METHODS: We compared the ECG-C rhythm interpretation to that of 2 expert overreaders in 2112 randomly selected standard 12-lead ECGs.

RESULTS: The ECG-C correctly interpreted the rhythm in 1858 and incorrectly identified the rhythm in 254 (overall accuracy, 88.0%). Sinus rhythm was correctly interpreted in 95.0% of the ECGs (1666/1753) with this rhythm, whereas nonsinus rhythms were correctly interpreted with an accuracy of only 53.5% (192/359) (P < .0001). The ECG-C interpreted sinus rhythm with a sensitivity of 95% (confidence interval, 93.8-96.7), specificity of 66.3%, and positive predictive value of 93.2%. The ECG-C interpreted nonsinus rhythms with a sensitivity of 72%, (confidence interval, 68.7-73.7), a specificity of 93%, and a positive predictive value of 59.3%. Of the 254 ECGs that had incorrect rhythm interpretation, additional major errors were noted in 137 (54%).

CONCLUSIONS: The ECG-C demonstrates frequent errors in the interpretation of nonsinus rhythms. In addition, incorrect rhythm interpretation by the ECG-C was frequently further compounded by additional major inaccuracies. Expert overreading of the ECG remains important in clinical settings with a high percentage of nonsinus rhythms.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
17531257
×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"