Impedance cardiography: a useful and reliable tool in optimization of cardiac resynchronization devices

Konstantin M Heinroth, Marcel Elster, Sebastian Nuding, Frithjof Schlegel, Arnd Christoph, Justin Carter, Michael Buerke, Karl Werdan
Europace: European Pacing, Arrhythmias, and Cardiac Electrophysiology 2007, 9 (9): 744-50
17496288

AIMS: Optimizing cardiac resynchronization therapy (CRT) devices has become more complex since modification of both atrioventricular (AV) and interventricular (VV) stimulation intervals has become possible. The current paper presents data from the routine use of impedance cardiography (IC)-based cardiac output (CO) measurements to guide the optimization of AV- and VV-interval timing of CRT devices.

METHODS AND RESULTS: Forty-six patients with heart failure (left ventricular ejection fraction <35%, New York Heart Association (NYHA) III-IV) and left bundle branch block (>130 ms) in sinus rhythm were evaluated 3-5 days after implantation of a CRT device by means of IC. CO was measured without pacing and with biventricular pacing using a standard protocol of VV- and AV-interval modification from -60 to +60 ms and 80 to 140 ms, respectively, in 20 ms steps. Mean CO without pacing was 3.66 +/- 0.85 L/min and significantly increased to 4.40 +/- 1.1 L/min (P<0.05) with simultaneous biventricular pacing and an AV interval of 120 ms. 'Optimizing' both VV and AV intervals further increased CO to 4.86 +/- 1.1 L/min (P<0.05). Maximum CO was measured in most patients with left ventricular pre-excitation. The proportion of 'non-responders' to CRT was reduced by 56% following AV- and VV-interval modification using IC guidance.

CONCLUSION: Modification of both AV and VV intervals in patients with a CRT device significantly improves CO compared with standard simultaneous biventricular pacing and no pacing. IC is a useful non-invasive technique for guiding this modification. Marked variability of optimal AV and VV intervals between patients requires optimization of these intervals for each patient individually.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
17496288
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"