Cone-beam computed tomography for on-line image guidance of lung stereotactic radiotherapy: localization, verification, and intrafraction tumor position

Thomas G Purdie, Jean-Pierre Bissonnette, Kevin Franks, Andrea Bezjak, David Payne, Fanny Sie, Michael B Sharpe, David A Jaffray
International Journal of Radiation Oncology, Biology, Physics 2007 May 1, 68 (1): 243-52
17331671

PURPOSE: Cone-beam computed tomography (CBCT) in-room imaging allows accurate inter- and intrafraction target localization in stereotactic body radiotherapy of lung tumors.

METHODS AND MATERIALS: Image-guided stereotactic body radiotherapy was performed in 28 patients (89 fractions) with medically inoperable Stage T1-T2 non-small-cell lung carcinoma. The targets from the CBCT and planning data set (helical or four-dimensional CT) were matched on-line to determine the couch shift required for target localization. Matching based on the bony anatomy was also performed retrospectively. Verification of target localization was done using either megavoltage portal imaging or CBCT imaging; repeat CBCT imaging was used to assess the intrafraction tumor position.

RESULTS: The mean three-dimensional tumor motion for patients with upper lesions (n = 21) and mid-lobe or lower lobe lesions (n = 7) was 4.2 and 6.7 mm, respectively. The mean difference between the target and bony anatomy matching using CBCT was 6.8 mm (SD, 4.9, maximum, 30.3); the difference exceeded 13.9 mm in 10% of the treatment fractions. The mean residual error after target localization using CBCT imaging was 1.9 mm (SD, 1.1, maximum, 4.4). The mean intrafraction tumor deviation was significantly greater (5.3 mm vs. 2.2 mm) when the interval between localization and repeat CBCT imaging (n = 8) exceeded 34 min.

CONCLUSION: In-room volumetric imaging, such as CBCT, is essential for target localization accuracy in lung stereotactic body radiotherapy. Imaging that relies on bony anatomy as a surrogate of the target may provide erroneous results in both localization and verification.

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Available on the App Store

Available on the Play Store
Remove bar
Read by QxMD icon Read
17331671
×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"