Comparison of localization performance with implanted fiducial markers and cone-beam computed tomography for on-line image-guided radiotherapy of the prostate

Douglas J Moseley, Elizabeth A White, Kirsty L Wiltshire, Tara Rosewall, Michael B Sharpe, Jeffrey H Siewerdsen, Jean-Pierre Bissonnette, Mary Gospodarowicz, Padraig Warde, Charles N Catton, David A Jaffray
International Journal of Radiation Oncology, Biology, Physics 2007 March 1, 67 (3): 942-53
17293243

PURPOSE: The aim of this work was to assess the accuracy of kilovoltage (kV) cone-beam computed tomography (CBCT)-based setup corrections as compared with orthogonal megavoltage (MV) portal image-based corrections for patients undergoing external-beam radiotherapy of the prostate.

METHODS AND MATERIALS: Daily cone-beam CT volumetric images were acquired after setup for patients with three intraprostatic fiducial markers. The estimated couch shifts were compared retrospectively to patient adjustments based on two orthogonal MV portal images (the current clinical standard of care in our institution). The CBCT soft-tissue based shifts were also estimated by digitally removing the gold markers in each projection to suppress the artifacts in the reconstructed volumes. A total of 256 volumetric images for 15 patients were analyzed.

RESULTS: The Pearson coefficient of correlation for the patient position shifts using fiducial markers in MV vs. kV was (R2 = 0.95, 0.84, 0.81) in the left-right (LR), anterior-posterior (AP), and superior-inferior (SI) directions, respectively. The correlation using soft-tissue matching was as follows: R2 = 0.90, 0.49, 0.51 in the LR, AP and SI directions. A Bland-Altman analysis showed no significant trends in the data. The percentage of shifts within a +/-3-mm tolerance (the clinical action level) was 99.7%, 95.5%, 91.3% for fiducial marker matching and 99.5%, 70.3%, 78.4% for soft-tissue matching.

CONCLUSIONS: Cone-beam CT is an accurate and precise tool for image guidance. It provides an equivalent means of patient setup correction for prostate patients with implanted gold fiducial markers. Use of the additional information provided by the visualization of soft-tissue structures is an active area of research.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
17293243
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"