We have located links that may give you full text access.
JOURNAL ARTICLE
RESEARCH SUPPORT, NON-U.S. GOV'T
Unconstrained tripolar hip implants: effect on hip stability.
Clinical Orthopaedics and related Research 2007 Februrary
Tripolar implants were developed to treat unstable total hip arthroplasties. However, there is limited confirmation that they achieve this purpose despite their increasing use. Because they have a larger effective head size, these implants are expected to increase range of motion to impingement and improve stability in situations at risk for impingement compared with conventional implants. We assessed the range of motion to impingement using a tripolar implant mounted to an automated hip simulator using 22.2-mm and 28-mm femoral head sizes. The 22 and 28-mm tripolar implants provided increases of 30.5 degrees in flexion, 15.4 degrees in adduction, and 22.4 degrees in external rotation compared with the conventional 22.2-mm femoral head diameter implant. At the critical position of 90 degrees hip flexion, there was an increase of 45.2 degrees in internal rotation. At 0 degrees and 30 degrees external rotation, extension increases were 18.8 degrees and 7.8 degrees, respectively. Bony impingement was the limiting factor. Tripolar implants increased the arc of motion before impingement in positions at risk for dislocation and are expected to provide greater stability.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app