Preoxygenation with the Mapleson D system requires higher oxygen flows than Mapleson A or circle systems.
Canadian Journal of Anaesthesia 2007 Februrary
PURPOSE: This study investigates the efficacy of preoxygenation with Mapleson A and Mapleson D breathing systems vs the circle system with CO2 absorber.
METHODS: Thirteen healthy volunteers underwent tidal volume breathing for three minutes via facemask using Mapleson A, Mapleson D breathing systems or the circle system with CO2 absorber while breathing 100% O2 at flow rates of 5 L.min-1 and 10 L.min-1. Each volunteer acted as his/her own control by going through each of six preoxygenation protocols in random order. Fractional end-tidal O2 concentration (FETO2) was measured at 30-sec intervals. The results were compared among the three anesthesia systems at the two fresh gas flow rates.
RESULTS: At a fresh gas flow rate of 5 L.min-1, the Mapleson A and circle systems achieved F(ETO2) values of 90.8+/-1.4% and 90.0+/-1.1%, respectively, compared with the lower F(ETO2) (81.5+/-6.3%, P<0.05), achieved with the Mapleson D system. When breathing O2 at 10 L.min-1, the F(ETO2) values after three minutes were similar with the Mapleson A, circle, and Mapleson D breathing systems (91.8+/-2.3%, 91.2+/-1.7%, 90.6+/-2.7%, respectively).
CONCLUSION: When using the Mapleson A and the circle systems for preoxygenation, an oxygen flow rate of 5 L.min-1 can adequately preoxygenate the patient within three minutes, while an oxygen flow of 10 L.min-1 is required to achieve a similar fractional end-tidal O2 concentration with the Mapleson D system.
METHODS: Thirteen healthy volunteers underwent tidal volume breathing for three minutes via facemask using Mapleson A, Mapleson D breathing systems or the circle system with CO2 absorber while breathing 100% O2 at flow rates of 5 L.min-1 and 10 L.min-1. Each volunteer acted as his/her own control by going through each of six preoxygenation protocols in random order. Fractional end-tidal O2 concentration (FETO2) was measured at 30-sec intervals. The results were compared among the three anesthesia systems at the two fresh gas flow rates.
RESULTS: At a fresh gas flow rate of 5 L.min-1, the Mapleson A and circle systems achieved F(ETO2) values of 90.8+/-1.4% and 90.0+/-1.1%, respectively, compared with the lower F(ETO2) (81.5+/-6.3%, P<0.05), achieved with the Mapleson D system. When breathing O2 at 10 L.min-1, the F(ETO2) values after three minutes were similar with the Mapleson A, circle, and Mapleson D breathing systems (91.8+/-2.3%, 91.2+/-1.7%, 90.6+/-2.7%, respectively).
CONCLUSION: When using the Mapleson A and the circle systems for preoxygenation, an oxygen flow rate of 5 L.min-1 can adequately preoxygenate the patient within three minutes, while an oxygen flow of 10 L.min-1 is required to achieve a similar fractional end-tidal O2 concentration with the Mapleson D system.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app