A randomized prospective study comparing the Cobra Perilaryngeal Airway and Laryngeal Mask Airway-Classic during controlled ventilation for gynecological laparoscopy.
Anesthesia and Analgesia 2007 January
BACKGROUND: An increasing number of noninvasive, supraglottic airway devices are currently available. In this randomized single-blind study, we compared the Cobra Perilaryngeal Airway (CobraPLA) to the [Laryngeal Mask Airway (LMA)-Classic] during gynecological laparoscopy.
METHODS: Forty patients received either an LMA-Classic or a CobraPLA. Insertion, ventilation and removal characteristics were noted, as well as any throat morbidity.
RESULTS: Devices were similar for insertion characteristics, adverse events, and throat morbidity. Before pneumoperitoneum, peak airway pressures were 20.3 +/- 4.9 cm H2O in the LMA-Classic group versus 25.5 +/- 7.9 cm H2O in the CobraPLA group, P = 0.01. This difference was maintained during pneumoperitoneum; LMA-Classic (22.8 +/- 6.1 cm H2O) and CobraPLA (28.1 +/- 8.5 cm H2O), P = 0.04. Macroscopic blood occurred only on the CobraPLA, seen on 40% of the devices after removal, P = 0.001.
CONCLUSION: During gynecological laparoscopy, the CobraPLA provides similar insertion characteristics, but higher airway sealing pressures than the LMA-Classic. The usefulness of this finding requires further investigation.
METHODS: Forty patients received either an LMA-Classic or a CobraPLA. Insertion, ventilation and removal characteristics were noted, as well as any throat morbidity.
RESULTS: Devices were similar for insertion characteristics, adverse events, and throat morbidity. Before pneumoperitoneum, peak airway pressures were 20.3 +/- 4.9 cm H2O in the LMA-Classic group versus 25.5 +/- 7.9 cm H2O in the CobraPLA group, P = 0.01. This difference was maintained during pneumoperitoneum; LMA-Classic (22.8 +/- 6.1 cm H2O) and CobraPLA (28.1 +/- 8.5 cm H2O), P = 0.04. Macroscopic blood occurred only on the CobraPLA, seen on 40% of the devices after removal, P = 0.001.
CONCLUSION: During gynecological laparoscopy, the CobraPLA provides similar insertion characteristics, but higher airway sealing pressures than the LMA-Classic. The usefulness of this finding requires further investigation.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app