Validation of a new D-dimer microparticle enzyme immunoassay (AxSYM D-Dimer) in patients with suspected pulmonary embolism (PE).

OBJECTIVES: The aim of the study was to evaluate a new automated assay for D-dimer testing (AxSYM D-Dimer) based on microparticle enzyme-immunoassay technology by comparing it with three well established D-dimer assays.

PATIENTS AND METHODS: The performance of the new assay was evaluated in 280 plasma samples that were collected prospectively from out-patients included in a management study evaluating a decision based algorithm.

RESULTS: 58/280 patients (21%) had PE diagnosed by CT. Median values of AxSYM D-dimer in patients with PE were 3689 ng/mL (range 775-9000). Comparison analysis displayed excellent agreement with VIDAS (kappa=0.84) and Asserachrom (kappa=0.81) D-dimer assays. A strong correlation was found between AxSYM and the VIDAS (r=0.96) and Asserachrom (r=0.89) D-dimer assays. The highest cut-off value for AxSYM that yielded a sensitivity of 100% was 765 ng/mL with a specificity of 50%. At the cut-off level <500 ng/mL, the sensitivity and specificity of AxSYM D-dimer were 100% and 34%; VIDAS 100% and 42%; Asserachrom 100% and 40%; and STALiatest 100% and 37%, respectively. AxSYM D-dimer was negative in 75 patients (33.8%). None of these had PE at the initial work-up or VTE during the 3-month follow-up.

CONCLUSIONS: AxSYM D-dimer seems to be safe and effective in ruling out PE in out-patients. The cut-off level can be set at 500 to 750 ng/mL, at which the assay displays a performance that is comparable to that of the ELISA based assays. However, further studies are needed to confirm the safety of the assay and to determine the most optimal cut-off level in patients with venous thromboembolism.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app