We have located links that may give you full text access.
EVALUATION STUDIES
JOURNAL ARTICLE
Diagnosis of early coagulation abnormalities in trauma patients by rotation thrombelastography.
Journal of Thrombosis and Haemostasis : JTH 2007 Februrary
BACKGROUND: Reagent-supported thromboelastometry with the rotation thrombelastography (e.g. ROTEM) is a whole blood assay that evaluates the visco-elastic properties during blood clot formation and clot lysis. A hemostatic monitor capable of rapid and accurate detection of clinical coagulopathy within the resuscitation room could improve management of bleeding after trauma.
OBJECTIVES: The goals of this study were to establish whether ROTEM correlated with standard coagulation parameters to rapidly detect bleeding disorders and whether it can help to guide transfusion.
METHODS: Ninety trauma patients were included in the study. At admission, standard coagulation assays were performed and ROTEM parameters such as clot formation time (CFT) and clot amplitude (CA) were obtained at 15 min (CA(15)) with two activated tests (INTEM, EXTEM) and at 10 min (CA(10)) with a test analyzing specifically the fibrin component of coagulation (FIBTEM).
RESULTS: Trauma induced significant modifications of coagulation as assessed by standard assays and ROTEM. A significant correlation was found between prothrombin time (PT) and CA(15)-EXTEM (r = 0.66, P < 0.0001), between activated partial thromboplastin time and CFT-INTEM (r = 0.91, P < 0.0001), between fibrinogen level and CA(10)-FIBTEM (r = 0.85, P < 0.0001), and between platelet count and CA(15)-INTEM (r = 0.57, P < 0.0001). A cutoff value of CA(15)-EXTEM at 32 mm and CA(10)-FIBTEM at 5 mm presented a good sensitivity (87% and 91%) and specificity (100% and 85%) to detect a PT > 1.5 of control value and a fibrinogen less than 1 g L(-1), respectively.
CONCLUSIONS: ROTEM is a point-of-care device that rapidly detects systemic changes of in vivo coagulation in trauma patients, and it might be a helpful device in guiding transfusion.
OBJECTIVES: The goals of this study were to establish whether ROTEM correlated with standard coagulation parameters to rapidly detect bleeding disorders and whether it can help to guide transfusion.
METHODS: Ninety trauma patients were included in the study. At admission, standard coagulation assays were performed and ROTEM parameters such as clot formation time (CFT) and clot amplitude (CA) were obtained at 15 min (CA(15)) with two activated tests (INTEM, EXTEM) and at 10 min (CA(10)) with a test analyzing specifically the fibrin component of coagulation (FIBTEM).
RESULTS: Trauma induced significant modifications of coagulation as assessed by standard assays and ROTEM. A significant correlation was found between prothrombin time (PT) and CA(15)-EXTEM (r = 0.66, P < 0.0001), between activated partial thromboplastin time and CFT-INTEM (r = 0.91, P < 0.0001), between fibrinogen level and CA(10)-FIBTEM (r = 0.85, P < 0.0001), and between platelet count and CA(15)-INTEM (r = 0.57, P < 0.0001). A cutoff value of CA(15)-EXTEM at 32 mm and CA(10)-FIBTEM at 5 mm presented a good sensitivity (87% and 91%) and specificity (100% and 85%) to detect a PT > 1.5 of control value and a fibrinogen less than 1 g L(-1), respectively.
CONCLUSIONS: ROTEM is a point-of-care device that rapidly detects systemic changes of in vivo coagulation in trauma patients, and it might be a helpful device in guiding transfusion.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app