COMPARATIVE STUDY
JOURNAL ARTICLE
RETRACTED PUBLICATION
Add like
Add dislike
Add to saved papers

Noninvasive assessment of cardiac output using thoracic electrical bioimpedance in hemodynamically stable and unstable patients after cardiac surgery: a comparison with pulmonary artery thermodilution.

OBJECTIVE: To compare noninvasive cardiac output (CO)measurement obtained with a new thoracic electrical bioimpedance (TEB) device, using a proprietary modification of the impedance equation, with invasive measurement obtained via pulmonary artery thermodilution.

DESIGN: Prospective, observational study.

SETTING: Surgical intensive care unit (ICU) of a university-affiliated community hospital.

PATIENTS AND PARTICIPANTS: Seventy-four adult patients undergoing elective cardiac surgery with routine pulmonary artery catheter placement.

INTERVENTIONS: None.

MEASUREMENTS AND RESULTS: Simultaneous paired CO and cardiac index (CI) measurements by TEB and thermodilution were obtained in mechanically ventilated patients upon admission to the ICU. For analysis of CI data the patients were subdivided into a hemodynamically stable group and a hemodynamically unstable group. The groups were analyzed using linear regression and tests of bias and precision. We found a significant correlation between thermodilution and TEB (r = 0.83; n < 0.001), accompanied by a bias of -0.01 l/min/m(2) and a precision of +/-0.57 l/min/m(2) for all CI data pairs. Correlation, bias, and precision were not influenced by stratification of the data. The correlation coefficient, bias, and precision for CI were 0.86 (n< 0.001), 0.03 l/min/m(2), and +/-0.47 l/min/m(2) in hemodynamically stable patients and 0.79 (n< 0.001), 0.06 l/min/m(2), and +/-0.68 l/min/m(2) in hemodynamically unstable patients.

CONCLUSIONS: Our results demonstrate a close correlation and clinically acceptable agreement and precision between CO measurements obtained with impedance cardiography using a new algorithm to calculate CO from variations in TEB, and those obtained with the clinical standard of care, pulmonary artery thermodilution, in hemodynamically stable and unstable patients after cardiac surgery.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app