Journal Article
Research Support, Non-U.S. Gov't
Add like
Add dislike
Add to saved papers

Characterization and in vivo evaluation of sol-gel derived hydroxyapatite coatings on Ti6Al4V substrates.

Hydroxyapatite (HAP) based coatings on titanium alloy Ti6Al4V were attempted by sol-gel process. Precursor sol was dip-coated onto the substrates and the formation of HAP as the main phase was ensured after the sintering of coatings at 900 degrees C, as confirmed by the characterization studies involving XRD, FT-IR, Raman, and SEM-EDXS analysis. The biocompatibility of HAP-based coated titanium implants was tested by implantation studies in rabbit's tibia for a period of 8 weeks. Bone remodelling and inflammatory responses of the implants were studied during the implantation period. The better tissue-implant interaction of the coated implants with respect to the uncoated implants was observed after a defect healing period of 8 weeks. A coated implant with no evidence of rejection is a further advantage of the sol-gel derived HAP coating.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app