VersaStep trocar hernia rate in unclosed fascial defects in bariatric patients.
Surgical Endoscopy 2006 October
OBJECTIVE: Use of the VersaStep trocar system (US Surgical, Norwalk, CT) has the perceived advantage of minimal trocar-related hernias in patients undergoing Roux-en-Y gastric bypass surgery (RYGB). We performed a retrospective review of our last 747 consecutive operative procedures using these trocars.
METHODS AND PROCEDURES: The patient population was 747 consecutive patients who underwent laparoscopic RYGB at Duke University Health System Weight Loss Surgery Center from January 2002 through April 2005. A total of 3735 radially expanded trocar sites were used. VersaStep trocars were used in all cases. The port configuration included one supraumbilical Hasson port, two 12-mm ports, and three 5-mm ports. The Hasson port was closed with a figure-of-eight number 1 Polysorb suture. All other trocar sites had no fascial closure. Intestinal anastomoses were created with a linear stapler in all of the laparoscopic cases, with hand suturing of the residual enterotomy. The fascial incisions were therefore not extended to accommodate an EEA stapler. The charts were reviewed for occurrence of subsequent trocar site hernias.
RESULTS: There were no hernias at any of the VersaStep trocar sites-an incidence of 0%. There were nine incisional hernias at the Hasson port site which later required surgical repair-an incidence of 1.20%.
CONCLUSIONS: There were no hernias detected at any of the 1494 12-mm or 2241 5-mm VersaStep trocar sites, despite lack of suture closure. At the Hasson port site, there was a hernia incidence of 1.20%. In the bariatric RYGB population, routine suture closure of the fascia or muscle is not necessary when using radially expanding VersaStep trocars.
METHODS AND PROCEDURES: The patient population was 747 consecutive patients who underwent laparoscopic RYGB at Duke University Health System Weight Loss Surgery Center from January 2002 through April 2005. A total of 3735 radially expanded trocar sites were used. VersaStep trocars were used in all cases. The port configuration included one supraumbilical Hasson port, two 12-mm ports, and three 5-mm ports. The Hasson port was closed with a figure-of-eight number 1 Polysorb suture. All other trocar sites had no fascial closure. Intestinal anastomoses were created with a linear stapler in all of the laparoscopic cases, with hand suturing of the residual enterotomy. The fascial incisions were therefore not extended to accommodate an EEA stapler. The charts were reviewed for occurrence of subsequent trocar site hernias.
RESULTS: There were no hernias at any of the VersaStep trocar sites-an incidence of 0%. There were nine incisional hernias at the Hasson port site which later required surgical repair-an incidence of 1.20%.
CONCLUSIONS: There were no hernias detected at any of the 1494 12-mm or 2241 5-mm VersaStep trocar sites, despite lack of suture closure. At the Hasson port site, there was a hernia incidence of 1.20%. In the bariatric RYGB population, routine suture closure of the fascia or muscle is not necessary when using radially expanding VersaStep trocars.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app