A comparison of polysomnography and the SleepStrip in the diagnosis of OSA.
Otolaryngology - Head and Neck Surgery 2006 August
OBJECTIVE: To investigate the role of a portable screening device (SleepStrip) in the diagnosis of obstructive sleep apnea (OSA).
METHODS AND MATERIALS: Prospective, nonrandomized double-blinded single cohort study at an academic health center. Patients with suspected OSA scheduled for an attended overnight Level I polysomnogram (PSG) and who consented to participate in the study wore the SleepStrip device at home the night after the PSG. The apnea-hypopnea index (AHI) determined by PSG was compared with the results of the SleepStrip recording.
RESULTS: Thirty-seven patients with a mean age of 52.1 +/- 12.2 years and mean body mass index of 35.7 +/- 5.2 participated in the study. The overall agreement between the AHI and the SleepStrip results using Cohen's Kappa value was 0.139 (P = 0.19). The sensitivity and specificity of the SleepStrip for diagnosing severe OSA when the AHI was >40 were 33.3% and 95% (P = 0.05). When the AHI was >25, the SleepStrip sensitivity and specificity were 43.8% and 81.3% (P = 0.26). The sensitivity and specificity of the SleepStrip for diagnosing OSA in patients with an AHI >15 were 54.6% and 70%, respectively (P = 0.26).
CONCLUSION: The SleepStrip has a low correlation with the AHI as measured by PSG. Further studies are needed before this device can be recommended as a screening tool for the diagnosis of OSA.
EBM RATING: B-2b.
METHODS AND MATERIALS: Prospective, nonrandomized double-blinded single cohort study at an academic health center. Patients with suspected OSA scheduled for an attended overnight Level I polysomnogram (PSG) and who consented to participate in the study wore the SleepStrip device at home the night after the PSG. The apnea-hypopnea index (AHI) determined by PSG was compared with the results of the SleepStrip recording.
RESULTS: Thirty-seven patients with a mean age of 52.1 +/- 12.2 years and mean body mass index of 35.7 +/- 5.2 participated in the study. The overall agreement between the AHI and the SleepStrip results using Cohen's Kappa value was 0.139 (P = 0.19). The sensitivity and specificity of the SleepStrip for diagnosing severe OSA when the AHI was >40 were 33.3% and 95% (P = 0.05). When the AHI was >25, the SleepStrip sensitivity and specificity were 43.8% and 81.3% (P = 0.26). The sensitivity and specificity of the SleepStrip for diagnosing OSA in patients with an AHI >15 were 54.6% and 70%, respectively (P = 0.26).
CONCLUSION: The SleepStrip has a low correlation with the AHI as measured by PSG. Further studies are needed before this device can be recommended as a screening tool for the diagnosis of OSA.
EBM RATING: B-2b.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app