Use of end-tidal partial pressure of carbon dioxide to predict arterial partial pressure of carbon dioxide in harp seals during isoflurane-induced anesthesia.
American Journal of Veterinary Research 2006 July
OBJECTIVE: To evaluate the relationship between end-tidal partial pressure of CO(2) (ETCO(2)) and PaCO(2) in isoflurane-anesthetized harp seals.
ANIMALS: Three 5-month-old 25- to 47-kg harp seals (Phoca groenlandica).
PROCEDURES: PaCO(2) was determined in serial arterial samples from isoflurane-anesthetized seals and compared with concomitant ETCO(2) measured with a side-stream microstream capnograph. Twenty-four paired samples were subjected to linear regression analysis and the Bland-Altman method for assessment of clinical suitability of the 2 methods (ie, PaCO(2) and ETCO(2) determinations). The influence of ventilation rate per minute (VR) on the ETCO(2) to PaCO(2) difference (P[ET-a] CO(2)) was examined graphically.
RESULTS: The correlation coefficient between the 2 measurements was 0.94. The level of agreement between ETCO(2) and PaCO(2) varied considerably. Values of ETCO(2) obtained with a VR of < 5 underestimated PaCO(2) to a greater degree (mean bias, -4.01 mm Hg) and had wider limits of agreement of -13.10 to 5.07 mm Hg (-4.01 mm Hg +/- 1.96 SD), compared with a VR of > or = 5 (mean bias, -2.24 mm Hg; limits of agreement, -7.79 to 3.30 mm Hg).
CONCLUSIONS AND CLINICAL RELEVANCE: These results indicate that a microstream sidestream capnograph provides a noninvasive, sufficiently accurate estimation of PaCO(2) with intermittent positive ventilation at a VR > or = 5 in anesthetized harp seals.
ANIMALS: Three 5-month-old 25- to 47-kg harp seals (Phoca groenlandica).
PROCEDURES: PaCO(2) was determined in serial arterial samples from isoflurane-anesthetized seals and compared with concomitant ETCO(2) measured with a side-stream microstream capnograph. Twenty-four paired samples were subjected to linear regression analysis and the Bland-Altman method for assessment of clinical suitability of the 2 methods (ie, PaCO(2) and ETCO(2) determinations). The influence of ventilation rate per minute (VR) on the ETCO(2) to PaCO(2) difference (P[ET-a] CO(2)) was examined graphically.
RESULTS: The correlation coefficient between the 2 measurements was 0.94. The level of agreement between ETCO(2) and PaCO(2) varied considerably. Values of ETCO(2) obtained with a VR of < 5 underestimated PaCO(2) to a greater degree (mean bias, -4.01 mm Hg) and had wider limits of agreement of -13.10 to 5.07 mm Hg (-4.01 mm Hg +/- 1.96 SD), compared with a VR of > or = 5 (mean bias, -2.24 mm Hg; limits of agreement, -7.79 to 3.30 mm Hg).
CONCLUSIONS AND CLINICAL RELEVANCE: These results indicate that a microstream sidestream capnograph provides a noninvasive, sufficiently accurate estimation of PaCO(2) with intermittent positive ventilation at a VR > or = 5 in anesthetized harp seals.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app