COMPARATIVE STUDY
JOURNAL ARTICLE

Comparison between the Hybrid Capture II Test and an SPF1/GP6+ PCR-based assay for detection of human papillomavirus DNA in cervical swab samples

Shang-Lang Huang, Angel Chao, Swei Hsueh, Fang-Yu Chao, Chu-Chun Huang, Jung-Erh Yang, Ching-Yu Lin, Chiu-Cho Yan, Hung-Hsueh Chou, Kuan-Gen Huang, Huei-Jean Huang, Tzu-I Wu, Mao-Jung Tseng, Jian-Tai Qiu, Cheng-Tao Lin, Ting-Chang Chang, Chyong-Huey Lai
Journal of Clinical Microbiology 2006, 44 (5): 1733-9
16672400
We compared the efficacy of human papillomavirus (HPV) DNA detection between a PCR-based genechip (Easychip HPV Blot [hereafter referred to as HPV Blot]; King Car, Taiwan) method and Hybrid Capture II (HCII; Digene, Gaithersburg, MD) in women with previous normal (n = 146) or abnormal (> or =atypical squamous cells of undetermined significance [ASCUS] [n = 208]) cytology. A total of 354 cervical swab samples were collected for HPV DNA assay by both HCII and SPF1/GP6+ PCR followed by HPV Blot tests. Colposcopy-directed biopsy was performed if clinically indicated. Of the 354 samples, HPV-positive rates by these two methods (HCII and HPV Blot) were 12.6% and 18.2% in 143 normal samples, 36.2% and 45.7% in 105 ASCUS samples, 57.4% and 57.4% in 94 low-grade squamous intraepithelial lesion samples, and 83.3% and 75.0% in 12 high-grade squamous intraepithelial lesion samples, respectively. The concordance of HPV Blot and HCII was 80.8% (286/354), and the agreement between the methods (kappa value, 0.68) was substantial. Discrepancies were further investigated by at least one of the following three methods: direct sequencing, type-specific PCR, and HPV Blot genotyping of cervical biopsy tissue. In the 15 HCII-positive samples, HPV Blot detected only non-HCII HPV genotypes; results of further verification methods were consistent with the latter test in the 15 samples. Of the 20 samples with HCII-negative and HPV Blot-positive results, 18 were found to contain the 13 HCII high-risk genotypes by verification methods. In only 16.7% (3/18) of the HCII-positive but HPV Blot-negative samples, further studies detected the 13 HCII genotypes. We conclude that HPV Blot seemed comparable to HCII for detection of HPV DNA in cervical swab samples.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
16672400
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"