Effects of vascular perfusion on coagulation size in radiofrequency ablation of ex vivo perfused bovine livers.

OBJECTIVES: A standardized perfused ex vivo bovine liver model was used to evaluate the effect of organ perfusion on coagulation size and energy deposition during radiofrequency ablation (RFA) procedures.

MATERIALS AND METHODS: Bovine livers were perfused in a tank after rinsing the prepared liver vessels with anticoagulants. Tyrode's solution, oxygenated and heated to 36.5 degrees C, was used as perfusion medium. A flow and pressure controlled pump regulated Portal vein circulation; a dialysis machine provided pulsatile arterial circulation. Impedance-guided radiofrequency ablations were performed with 4-cm LeVeen electrodes with and without underlying liver perfusion. Two-dimensional diameters (Dv, Dh) of each ablation area were measured after dissecting the livers.

RESULTS: In 4 bovine livers weighing 8.85 +/- 0.83 kg per organ (min, 7.7 kg; max, 9.7 kg) altogether 40 RF ablations were performed. A total of 20 ablations were generated with underlying liver perfusion (group 1) and 20 ablations with no liver perfusion (group 2). In group 1, Dv was 28.4 +/- 5.3 mm, Dh 38.6 +/- 7.8 mm, and energy deposition 36.9 +/- 18.0 kJ. The 20 ablation areas generated without liver perfusion displayed statistically significant differences, with Dv being 35.7 +/- 6.5 mm (P = 0.001), Dh 49.5 +/- 9.4 mm (P = 0.001), and energy deposition 25.5 +/- 13.0 kJ (P = 0.018).

CONCLUSION: The model reproduced the cooling effect of perfused tissue during RFA. The ablation areas produced under perfusion conditions had smaller diameters despite longer exposure times and higher energy deposition.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app