In vivo porcine liver radiofrequency ablation with simultaneous MR temperature imaging.
Journal of Magnetic Resonance Imaging : JMRI 2006 April
PURPOSE: To demonstrate in vivo MR-guided temperature mapping during radiofrequency (RF) ablation of the liver with a commercially available RF generator modified to allow simultaneous RF treatment and MRI.
MATERIALS AND METHODS: A commercial RF generator was modified using passive filtering to allow the continuous application of the treatment current during MRI studies. A total of six ablations were performed with the device in vivo in three porcine livers, and imaging was concurrently performed using one of two different temperature mapping strategies.
RESULTS: MR images acquired during RF ablation demonstrated no noticeable interference from the RF ablation device, which was operated at clinically relevant power levels. Temperature maps showed areas of heating that were consistent with the dimensions of the RF ablation probe, with some asymmetry (likely depending on the orientation of the probe and heat propagation effects), and some differences in heating-spot area stability depending on the specific temperature mapping strategy used. Lesions were visualized on post-ablation imaging and sectioning.
CONCLUSION: The feasibility of performing RF ablation with a modified commercial RF generator simultaneously with MRI was demonstrated. Interference-free MR temperature maps were produced with both variable respiratory motion and mechanical ventilation, and showed the extent of heating as the ablation progressed.
MATERIALS AND METHODS: A commercial RF generator was modified using passive filtering to allow the continuous application of the treatment current during MRI studies. A total of six ablations were performed with the device in vivo in three porcine livers, and imaging was concurrently performed using one of two different temperature mapping strategies.
RESULTS: MR images acquired during RF ablation demonstrated no noticeable interference from the RF ablation device, which was operated at clinically relevant power levels. Temperature maps showed areas of heating that were consistent with the dimensions of the RF ablation probe, with some asymmetry (likely depending on the orientation of the probe and heat propagation effects), and some differences in heating-spot area stability depending on the specific temperature mapping strategy used. Lesions were visualized on post-ablation imaging and sectioning.
CONCLUSION: The feasibility of performing RF ablation with a modified commercial RF generator simultaneously with MRI was demonstrated. Interference-free MR temperature maps were produced with both variable respiratory motion and mechanical ventilation, and showed the extent of heating as the ablation progressed.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app