Simultaneous detection of antibody binding and cytotoxicity in flow cytometry crossmatch for renal transplantation

Dong Il Won, Hee Du Jeong, Yong Lim Kim, Jang Soo Suh
Cytometry. Part B, Clinical Cytometry 2006, 70 (2): 82-90
16470535

BACKGROUND: The anti-HLA antibody can be detected using either a complement-dependent lymphocytotoxicity (CDC) assay or a flow cytometry crossmatch (FCXM) in renal transplantation. Discordant results are often obtained because the two methods detect different reaction phases between the donor lymphocytes and the recipient sera. This study was intended to confirm that antibody binding and cytotoxicity to the lymphocytes can be detected simultaneously in a single FCXM assay, cytotoxic FCXM.

METHODS: In the cytotoxic FCXM, the antibody binding to the lymphocytes was measured using anti-IgG-FITC, and the cytotoxicity using 7-aminoactinomycin D (7-AAD) after adding complement. For an evaluation of two test parameters, the cytotoxicity test moiety (dead-cell percentage) was compared with the anti-human globulin (AHG)-CDC, and the antibody-binding test moiety (sample/control fluorescence ratio) with the conventional FCXM in 77 positive and 30 negative crossmatches.

RESULTS: In the cytotoxic FCXM, both antibody binding and cytotoxicity could be assessed in a single anti-IgG-FITC/7-AAD plot. Regarding the correlation between the presence of HLA antibodies and the test result, the cytotoxicity parameter (r = 0.55) appeared to be more suitable than that of the AHG-CDC (r = 0.50) but the antibody-binding parameter (r = 0.83) was worse than that of the conventional FCXM (r = 0.93). The sensitivity of both parameters of the cytotoxic FCXM was not significantly different from each conventional counterpart (P = 0.33 and P = 0.22, respectively).

CONCLUSIONS: The simultaneous detection of Ab binding and cytotoxicity was possible by the cytotoxic FCXM with the test efficiencies similar to the conventional counterparts. If this new assay is improved through the further studies to optimize the critical assay variables, this may be used as an alternative to the conventional assays to acquire more information on the characteristics of the recipient's HLA alloantibodies.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
16470535
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"