Effect of suture material on gap formation and failure in type 1 FDP avulsion repairs in a cadaver model.
Clinical Biomechanics 2006 June
BACKGROUND: An in vitro cyclical testing simulating a passive mobilisation protocol was used to compare repair of flexor digitorum profundus tendon with modified-Bunnell two-strand pullout technique using a monofilament (Prolene), braided polyester (Ethibond) and a synthetic polyfilament ensheathed by caprolactan (Supramid) sutures.
METHODS: Eighteen fresh-frozen cadaveric fingers were randomly divided into three repair groups (n = 6); modified-Bunnell technique with 3/0 Prolene, Ethibond or Supramid. After repair, specimens were cyclically loaded from 2 to 15N at 5N/s, for a total of 500 cycles. Gap formation at the tendon-bone interface was assessed every 100 cycles. Samples were tested to failure at the completion of 500 cycles.
FINDINGS: All sutures held in all specimens during cyclic testing. The gap formation after 500 cycles was greatest with Prolene suture (6.8 mm, SD 1.2) followed by Supramid suture (4.0 mm, SD 1.1) and Ethibond suture (1.7 mm, SD 1.7) (P < 0.05). Repairs with Supramid displayed higher failure load (52.7 N, SD 5.5) as compared to Prolene (37.6N, SD 4.7) (P = 0.001) but not compared to Ethibond (44.9 N, SD 7.1). The failure loads between Prolene and Ethibond did not differ (P = 0.130).
INTERPRETATION: Gap formation with Ethibond was significantly lower compared to Supramid and Prolene. The four strand nature of the Supramid repair was superior to Prolene but did not differ compared to Ethibond with respect to failure load. Prolene is the least favourable suture when considering gap formation and failure load, while Ethibond is the most favourable.
METHODS: Eighteen fresh-frozen cadaveric fingers were randomly divided into three repair groups (n = 6); modified-Bunnell technique with 3/0 Prolene, Ethibond or Supramid. After repair, specimens were cyclically loaded from 2 to 15N at 5N/s, for a total of 500 cycles. Gap formation at the tendon-bone interface was assessed every 100 cycles. Samples were tested to failure at the completion of 500 cycles.
FINDINGS: All sutures held in all specimens during cyclic testing. The gap formation after 500 cycles was greatest with Prolene suture (6.8 mm, SD 1.2) followed by Supramid suture (4.0 mm, SD 1.1) and Ethibond suture (1.7 mm, SD 1.7) (P < 0.05). Repairs with Supramid displayed higher failure load (52.7 N, SD 5.5) as compared to Prolene (37.6N, SD 4.7) (P = 0.001) but not compared to Ethibond (44.9 N, SD 7.1). The failure loads between Prolene and Ethibond did not differ (P = 0.130).
INTERPRETATION: Gap formation with Ethibond was significantly lower compared to Supramid and Prolene. The four strand nature of the Supramid repair was superior to Prolene but did not differ compared to Ethibond with respect to failure load. Prolene is the least favourable suture when considering gap formation and failure load, while Ethibond is the most favourable.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app