A biomechanical and radiographic analysis of standard and intracortical suture anchors for arthroscopic rotator cuff repair.

Arthroscopy 2006 Februrary
PURPOSE: To compare the fixation strength and radiographic motion of an anchor designed for intracortical (IC) fixation (FT Anchor, Arthrex, Naples, FL) with that of standard anchors used for rotator cuff repair.

TYPE OF STUDY: In vitro human cadaveric biomechanical study.

METHODS: Four types of metallic suture anchors (8 per group) were randomly inserted into human cadaveric humeri using an IC anchor and 3 types of standard anchors. Anchors were inserted 45 degrees to the humeral head surface and 90 degrees to the rotator cuff line of action. Anchors were tested under physiologic loads for 500 cycles followed by a failure test. The number of cycles, failure mode, and failure load were recorded. Fluoroscopy was used to measure rotation and displacement of the anchor within the humeral head during testing. Data were analyzed using a 1-way analysis of variance with a correction for multiple comparisons.

RESULTS: There were no significant differences in anchor displacement or rotation measured by fluoroscopy after cyclic loading. Total construct displacement across anchors ranged from 4.9 to 7.8 mm, well beyond the 3-mm failure criterion reported in the literature. The IC anchor had a statistically significant greater failure load than the other devices. There was no significant difference in failure load between the other 3 anchors. The anchor had the greatest number of cycles to 3 mm of failure. This was not significantly different than the TwinFix anchor (Smith & Nephew, Andover, MA), but both values were significantly greater than both the Super Revo (Linvatec, Largo, FL) and Fastin RC (DePuy Mitek, Raynham, MA) anchors.

CONCLUSIONS: Anchor motion accounted for about one third of total displacement of the suture/anchor construct. IC fixation anchors performed well compared with standard anchors in human cadaveric bone.

CLINICAL RELEVANCE: Fluoroscopic imaging showed both rotation and displacement of the anchor within the humeral head which may contribute to early gap formation after rotator cuff repairs.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app