Evaluation of a point-of-care direct spectrophotometric method for measurement of total serum bilirubin in term and near-term neonates.
OBJECTIVE: To evaluate point-of-care (POC) measurement of total serum bilirubin (TSB) in the management of neonatal jaundice.
STUDY DESIGN: TSB was measured by a POC direct spectrophotometric bilirubin method (Unistat (U/TSB)) and a standard diazo clinical laboratory method (Olympus AU640E analyzer (diazo/TSB)). Agreement between U/TSB and diazo/TSB was assessed by correlation coefficient and Bland-Altman analysis. Transcutaneous bilirubin (TcB) was measured using JM-103 (JM).
RESULTS: Correlation between U/TSB and diazo/TSB was 0.99 (n = 120). Maximum difference (U/TSB minus diazo/TSB) was -2.9 mg/dl, and 79% were +/-1 mg/dl; the average difference was -0.37+/-0.70 mg/dl and the average absolute difference was 0.60+/-0.52 mg/dl. Median time to determine U/TSB was 5 min. Correlation between U/TSB and JM was 0.92 (n = 113). Maximum difference (U/TSB minus JM) was 6.3 mg/dl, and 45% were +/-1 mg/dl; the average difference was 0.7+/-1.8 mg/dl and the average absolute difference was 1.4+/-1.2 mg/dl.
CONCLUSION: Measurement of TSB using Unistat provides excellent agreement with diazo/TSB and rapid turnaround time. This technique may provide reliable POC confirmation of TcB results that are above a screening cutoff value.
STUDY DESIGN: TSB was measured by a POC direct spectrophotometric bilirubin method (Unistat (U/TSB)) and a standard diazo clinical laboratory method (Olympus AU640E analyzer (diazo/TSB)). Agreement between U/TSB and diazo/TSB was assessed by correlation coefficient and Bland-Altman analysis. Transcutaneous bilirubin (TcB) was measured using JM-103 (JM).
RESULTS: Correlation between U/TSB and diazo/TSB was 0.99 (n = 120). Maximum difference (U/TSB minus diazo/TSB) was -2.9 mg/dl, and 79% were +/-1 mg/dl; the average difference was -0.37+/-0.70 mg/dl and the average absolute difference was 0.60+/-0.52 mg/dl. Median time to determine U/TSB was 5 min. Correlation between U/TSB and JM was 0.92 (n = 113). Maximum difference (U/TSB minus JM) was 6.3 mg/dl, and 45% were +/-1 mg/dl; the average difference was 0.7+/-1.8 mg/dl and the average absolute difference was 1.4+/-1.2 mg/dl.
CONCLUSION: Measurement of TSB using Unistat provides excellent agreement with diazo/TSB and rapid turnaround time. This technique may provide reliable POC confirmation of TcB results that are above a screening cutoff value.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app