Comparison of EASI-derived 12-lead electrocardiograms versus paramedic-acquired 12-lead electrocardiograms using Mason-Likar limb lead configuration in patients with chest pain.

INTRODUCTION: Monitoring or serial 12-lead electrocardiogram (ECG) recordings are the accepted requirement for prehospital data acquisition in patients with chest pain. The purpose of this study was to determine whether waveforms and clinical triage decision are similar in EASI-derived ECGs and paramedic-acquired 12-lead ECGs using Mason-Likar limb lead configuration when compared with standard 12-lead ECGs (stdECG).

METHOD: Twenty patients with chest pain had a prehospital 12-lead ECG recorded in the ambulance, and paramedic-applied electrodes retained in place at hospital arrival. An ECG technician applied standard precordial and EASI electrodes in their correct positions. Twelve-lead ECGs were obtained from the paramedic-applied electrodes, using their Mason-Likar limb lead configuration, and derived from the EASI leads for comparison with the stdECG. Three computer-measured QRS-T waveform parameters were considered, and differences in waveform measurement between EASI and stdECG (EASIDeltastdECG) versus differences in waveform measurements between paramedic Mason-Likar and stdECG (PMLDeltastdECG) were calculated. Two physicians determined whether the EASI-derived or the paramedic Mason-Likar ECG contained information that would change their clinical triage decision from that indicated by the stdECG.

RESULTS: EASIDeltastdECG and PMLDeltastdECG were identical in 28%, whereas EASIDeltastdECG was more than PMLDeltastdECG in 35%, and PMLDeltastdECG was accurate (both time) than EASIDeltastdECG in 37% (P = .62). The physicians were more likely to change the level of patient care based on the EASI-derived ECGs compared with the paramedic ECGs; however, this difference was not statistically significant (P = .27), but this may only be caused by the small study population.

CONCLUSIONS: There are similar differences from stdECG waveforms in EASI-derived ECGs and those acquired via paramedic-applied precordial electrodes using Mason-Likar limb lead configuration. Either method can be used as a substitute for monitoring, but neither should be considered equivalent to the stdECG for diagnostic purposes.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app