Forces at single point attached commissures (SPAC) in pericardial aortic valve prosthesis

Wolfgang A Goetz, Khee Hiang Lim, Romain Ibled, Nil Grousson, Sidney Le Hung Salgues, Joon Hock Yeo
European Journal of Cardio-thoracic Surgery 2006, 29 (2): 150-5
16386912

OBJECTIVE: New pericardial aortic bioprostheses (3F Therapeutics and temporarily stented autologous pericardial valve prosthesis) were developed recently. These valves are designed with commissures connected to the aortic wall at only three single points (single point attached commissures (SPAC)). The aim of this study was to investigate the forces acting on SPAC during varying pressure load.

METHODS: Aortic roots with diameters 19, 25, and 29 mm were made using silicone polymer. A bovine pericardial SPAC aortic valve prosthesis was constructed using a 3D-mold and was implanted in the silicone aortic root. The base of the valve was sutured onto the aortic annulus with 4-0 polypropylene running suture and each commissure was sutured to a miniaturized force transducer with only one 3-0 polypropylene U-stitch. Three silicon aortic roots of each size were pressurized up to 200 mmHg and forces on SPAC were measured.

RESULTS: All valves remained competent at a pressure of 200 mmHg. Recordings showed a linear correlation between applied pressure and forces measured at SPAC. At a pressure of 80 mmHg (equivalent to diastolic pressure), the forces were 0.44+/-0.22N, 1.15+/-0.18N, and 2.00+/-0.35N in annular diameters 19 mm, 25 mm, and 29 mm, respectively. It was observed, that the main forces were acting along the axial direction and not along the radial direction.

CONCLUSIONS: Forces on "single point attached commissures" in pericardial aortic valves were measured. These forces were acting mainly in axial direction and not in radial direction. This knowledge is important for the implantation technique of SPAC pericardial aortic valves.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
16386912
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"