JOURNAL ARTICLE
RANDOMIZED CONTROLLED TRIAL
RESEARCH SUPPORT, NON-U.S. GOV'T
Add like
Add dislike
Add to saved papers

Time course of left ventricular volumes in severe congestive heart failure patients treated by optimized AV sequential left ventricular pacing alone--a 3-dimensional echocardiographic study.

BACKGROUND: This study evaluates the acute and chronic resynchronizing effects of AV sequential left ventricular (LV) pacing on LV function in patients with impaired cardiac function and conduction disorders by 3-dimensional transesophageal echocardiography.

METHODS AND RESULTS: Twenty-nine patients with congestive heart failure, with LV ejection fraction (LVEF) < or = 30%, QRS duration > or = 120 milliseconds, and New York Heart Association Class II to IV, were implanted with a cardiac resynchronization device using an LV lead only, according to the invasively determined hemodynamic optimal pacing site and AV delay. Patients underwent 3-dimensional transesophageal echocardiography before randomization to treatment (baseline) and at 12-month follow-up (resynchronization--12 months). Three-dimensional volumes were acquired on resynchronization and during intermittent switch-off at intrinsic depolarization. The values of stroke volume were 43.2 +/- 13.3 (intrinsic-baseline), 51.7 +/- 17.4 (intrinsic--12 months), 57.2 +/- 15.6 (resynchronization-baseline), and 64.6 +/- 18.9 (resynchronization--12 months). Analysis of variance demonstrated a significant effect of resynchronization at different periods (P < .001) and a significant time effect (P < .05) for stroke volume. Similar results were observed with ejection fraction (LVEF). No effect was observed with LV end-diastolic volume, whereas a therapy effect with no time effect was observed with LV end-systolic volume.

CONCLUSIONS: A significant acute increase of LV stroke volume and LVEF was found by resynchronization by LV pacing alone. A continuous improvement of LV stroke volume and LVEF occurred with time of follow-up (reverse remodeling). The initial therapeutic effect persisted during 12-month follow-up independently of time of follow-up and QRS width. No significant decrease of LV end-diastolic size during chronic resynchronization was detected in contrast to previous studies with resynchronization by biventricular pacing.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app