OBJECTIVES: This study evaluated the pacing and sensing characteristics of a new porous-tipped steroid-eluting epicardial lead in a group of pediatric patients.

BACKGROUND: Pacing in children may be complicated by small patient size, patient growth and the prevalence of structural congenital heart disease in children requiring pacing. Epicardial pacing has been associated with a high incidence of problems with sensing and capture, prompting the use of transvenous endocardial pacing when possible. In some children, epicardial pacing may still be desirable because of small patient size, potential for caval obstruction, previous cardiac surgery limiting transvenous access to the heart, or the need to repair congenital heart disease at the time of pacemaker insertion.

METHODS: Twelve patients aged 3 weeks to 18 years underwent placement of 23 epicardial pacing leads (8 atrial, 15 ventricular). Pulse width thresholds, sensing thresholds and lead impedance were measured weekly for 6 weeks, then at 3, 6, 12 and 18 months after pacemaker implantation. The median duration of follow-up was 12 months.

RESULTS: Ventricular pulse width thresholds did not change over time, whereas atrial pulse width thresholds improved significantly. At 6 months, the mean pulse width threshold at 2.5 V for the atrial and ventricular leads was 0.10 +/- 0.03 and 0.19 +/- 0.09 ms, respectively. The thresholds were slightly lower at 12 and 18 months. At the most recent follow-up, all atrial leads sensed appropriately at 2.5 mV and all ventricular leads at 5 mV.

CONCLUSIONS: These encouraging early results suggest that steroid-eluting epicardial pacing leads may be an attractive option for children needing epicardial pacing. Their excellent pacing and sensing characteristics may allow reliable dual-chamber pacing in infants who are too small for transvenous pacing.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app