Evaluation of the AMPLATZER vascular plug for embolization of peripheral vascular malformations associated with congenital heart disease.
Catheterization and Cardiovascular Interventions 2006 January
OBJECTIVE: The purpose of this study is to evaluate the recently FDA-approved AMPLATZER Vascular Plug in the embolization of vascular lesions associated with congenital heart disease (CHD).
BACKGROUND: Fistulas and arteriovenous malformations have been occluded using various devices. The AMPLATZER Vascular Plug is a self-expandable, cylindrical device, used for embolization in the peripheral vasculature.
METHODS: A total of 84 vessels in 52 patients with CHD from 11 centers were occluded with 89 AMPLATZER Vascular Plugs, delivered through a coronary guide catheter in various vascular sites, including collaterals, pulmonary arterio-venous and coronary artery fistulas, transhepatic tracts, central shunts, patent ductus arteriosus (PDA), and excluded hepatic vein. Complete vessel occlusion was demonstrated within 10 min in 94% of patients.
RESULTS: There was no device embolization, vascular disruption, or procedure-related complication. One vascular plug implanted in a large type C PDA required surgical removal followed by PDA ligation, after 5 weeks from successful implant because of significant residual flow through the device.
CONCLUSIONS: The AMPLATZER Vascular Plug is an effective transcatheter occlusion device in the embolization of a wide variety of vascular lesions associated with CHD. Based on our early experience, caution should be used when considering the Vascular Plug as a closure device for large PDA.
BACKGROUND: Fistulas and arteriovenous malformations have been occluded using various devices. The AMPLATZER Vascular Plug is a self-expandable, cylindrical device, used for embolization in the peripheral vasculature.
METHODS: A total of 84 vessels in 52 patients with CHD from 11 centers were occluded with 89 AMPLATZER Vascular Plugs, delivered through a coronary guide catheter in various vascular sites, including collaterals, pulmonary arterio-venous and coronary artery fistulas, transhepatic tracts, central shunts, patent ductus arteriosus (PDA), and excluded hepatic vein. Complete vessel occlusion was demonstrated within 10 min in 94% of patients.
RESULTS: There was no device embolization, vascular disruption, or procedure-related complication. One vascular plug implanted in a large type C PDA required surgical removal followed by PDA ligation, after 5 weeks from successful implant because of significant residual flow through the device.
CONCLUSIONS: The AMPLATZER Vascular Plug is an effective transcatheter occlusion device in the embolization of a wide variety of vascular lesions associated with CHD. Based on our early experience, caution should be used when considering the Vascular Plug as a closure device for large PDA.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app