Zero filled partial fourier phase contrast MR imaging: in vitro and in vivo assessment.
Journal of Magnetic Resonance Imaging : JMRI 2006 January
PURPOSE: To validate partial Fourier phase contrast magnetic resonance (PC MR) with full number of excitation (NEX) PC MR measurements in vitro and in vivo.
MATERIALS AND METHODS: MR flow measurements were performed using a partial Fourier and a full NEX PC MR sequence in a flow phantom and in 10 popliteal and renal arteries of 10 different healthy volunteers. Average velocity, peak velocity, and flow results were calculated and compared with regression analysis.
RESULTS: Excellent correlations in average velocities (r = 0.99, P < 0.001), peak velocities (r = 0.99, P < 0.001), and flow rates (r = 0.98, P < 0.001) were demonstrated in vitro between the two different acquisitions. For the popliteal arteries there was excellent correlation between peak velocities for both acquisitions (r = 0.98, P < 0.0001); the correlation of average velocity measurements when using all data points in the cardiac cycle for all volunteers was 0.96 (P < 0.001). For the renal arteries the same comparison resulted in a good correlation for average velocity (0.93, P < 0.001) and peak velocity measurements (r = 0.91, P = 0.002), although the correlation coefficient for flow rates was 0.88 (P = 0.004). Blurring of the vessel margins was consistently observed on magnitude images acquired with the partial Fourier method, causing overestimation of the vessel area and some error in the flow measurements.
CONCLUSION: Partial Fourier PC MR is able to provide comparable average and peak velocity values when using 1 NEX PC MRI as a reference.
MATERIALS AND METHODS: MR flow measurements were performed using a partial Fourier and a full NEX PC MR sequence in a flow phantom and in 10 popliteal and renal arteries of 10 different healthy volunteers. Average velocity, peak velocity, and flow results were calculated and compared with regression analysis.
RESULTS: Excellent correlations in average velocities (r = 0.99, P < 0.001), peak velocities (r = 0.99, P < 0.001), and flow rates (r = 0.98, P < 0.001) were demonstrated in vitro between the two different acquisitions. For the popliteal arteries there was excellent correlation between peak velocities for both acquisitions (r = 0.98, P < 0.0001); the correlation of average velocity measurements when using all data points in the cardiac cycle for all volunteers was 0.96 (P < 0.001). For the renal arteries the same comparison resulted in a good correlation for average velocity (0.93, P < 0.001) and peak velocity measurements (r = 0.91, P = 0.002), although the correlation coefficient for flow rates was 0.88 (P = 0.004). Blurring of the vessel margins was consistently observed on magnitude images acquired with the partial Fourier method, causing overestimation of the vessel area and some error in the flow measurements.
CONCLUSION: Partial Fourier PC MR is able to provide comparable average and peak velocity values when using 1 NEX PC MRI as a reference.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app