Implantable cardioverter-defibrillators in patients with hypertrophic cardiomyopathy -- dilemmas and difficulties.

Kardiologia Polska 2005 October
INTRODUCTION: The implantation of a cardioverter-defibrillator (ICD) is an established method of sudden cardiac death (SCD) prevention. The value of ICD therapy in secondary prevention of SCD is unquestionable. Precise identification of high-risk patients and ICD use for primary prevention of SCD, especially in patients with hypertrophic cardiomyopathy (HCM), remain controversial. Problems include the high prevalence of complications associated with ICD implantation and optimal selection of ICDs.

AIM: To estimate the frequency and type of complications after ICD implantations in HCM patients in a long-term follow-up.

METHOD: The efficacy and safety of ICD therapy were estimated in 46 HCM patients with devices implanted for a secondary (n-18) or primary prevention (n-28) of SCD.

RESULTS: During the mean follow-up period of 28.2+/-26.1 months (from 2 to 68) appropriate ICD interventions occurred in 10 (55%) patients of the secondary prevention group and in 3 (10%) patients of the primary prevention group. Complications were documented in 15 (33%) patients. The most frequent were inappropriate ICD interventions recorded in 14 (30%) patients. The causes of these inappropriate ICD shocks were: T-wave oversensing (7 patients), atrial fibrillation with rapid ventricular rhythm (3 patients), lead failure (2 patients), and sinus tachycardia (2 patients). In two patients infections of the ICD pocket requiring removal of the system occurred. Displacement of the lead occurred in one patient. There were no significant differences in the prevalence of complications between the primary and secondary prevention groups or in the number of inappropriate interventions with respect to ICD type.

CONCLUSIONS: The high rate of appropriate ICD shocks provides proof of high ICD-based SCD prevention efficacy. There is a high rate of complications observed after ICD implantation with inappropriate interventions being the most frequent among them. This indicates that careful programming of the device as well as the use of a programme with T-wave oversensing prevention should be ensured.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app