Prediction of the recurrence of atrial fibrillation after successful cardioversion with P wave signal-averaged ECG.
Annals of Noninvasive Electrocardiology 2005 October
BACKGROUND: The recurrence of atrial fibrillation (AF) was often observed after cardioversion.
METHODS: In our study, a P wave triggered P wave signal-averaged ECG (P-SAECG) was performed on 118 consecutive patients 1 day after successful electrical cardioversion in order to evaluate the utility of this method to predict AF after cardioversion. We measured the filtered P wave duration (FPD) and the root mean square voltage of the last 20 ms of the P wave (RMS 20).
RESULTS: During a 1-year follow-up, a recurrence was observed in 57 patients (48%). Patients with recurrence of AF had a larger left atrial size (41.9 +/- 4.0 vs 39.3 +/- 3.1 mm, P < 0.0003), a longer FPD (139.6 +/- 16.0 vs 118.2 +/- 14.1 ms, P < 0.0001), and a lower RMS 20 (2.57 +/- 0.77 vs 3.90 +/- 0.99 microV, P < 0.0001). A cutoff point (COP) of FPD > or =126 ms and RMS 20 < or =3.1 microV could predict AF with a specificity of 77%, a sensitivity of 72%, a positive value of 75%, a negative predictive value of 75%, and an accuracy of 75%. A stepwise logistic regression analysis of variables identified COP (odds ratio 9.97; 95% CI, 4.10-24.24, P < 0.0001) as an independent predictor for recurrence.
CONCLUSIONS: We conclude that the probability of recurrence of AF after cardioversion could be predicted by P-SAECG. This method seems to be appropriate to demonstrate a delayed atrial conduction that might be a possible risk factor of reinitiation of AF.
METHODS: In our study, a P wave triggered P wave signal-averaged ECG (P-SAECG) was performed on 118 consecutive patients 1 day after successful electrical cardioversion in order to evaluate the utility of this method to predict AF after cardioversion. We measured the filtered P wave duration (FPD) and the root mean square voltage of the last 20 ms of the P wave (RMS 20).
RESULTS: During a 1-year follow-up, a recurrence was observed in 57 patients (48%). Patients with recurrence of AF had a larger left atrial size (41.9 +/- 4.0 vs 39.3 +/- 3.1 mm, P < 0.0003), a longer FPD (139.6 +/- 16.0 vs 118.2 +/- 14.1 ms, P < 0.0001), and a lower RMS 20 (2.57 +/- 0.77 vs 3.90 +/- 0.99 microV, P < 0.0001). A cutoff point (COP) of FPD > or =126 ms and RMS 20 < or =3.1 microV could predict AF with a specificity of 77%, a sensitivity of 72%, a positive value of 75%, a negative predictive value of 75%, and an accuracy of 75%. A stepwise logistic regression analysis of variables identified COP (odds ratio 9.97; 95% CI, 4.10-24.24, P < 0.0001) as an independent predictor for recurrence.
CONCLUSIONS: We conclude that the probability of recurrence of AF after cardioversion could be predicted by P-SAECG. This method seems to be appropriate to demonstrate a delayed atrial conduction that might be a possible risk factor of reinitiation of AF.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app